CALIBRATION PROCEDURE
FD-11603
8-Channel, ±10 V Voltage Input Device for FieldDAQ
This document contains information for calibrating the FD-11603. For more information about
calibration, visit
Contents
Software Requirements............................................................................................................. 1
Documentation.......................................................................................................................... 2
Test Equipment..........................................................................................................................2
Test Conditions..........................................................................................................................3
Calibration Procedure............................................................................................................... 3
Initial Setup....................................................................................................................... 4
Verification........................................................................................................................4
Adjustment........................................................................................................................ 7
EEPROM Update..............................................................................................................8
Reverification....................................................................................................................8
Accuracy Under Calibration Conditions...................................................................................8
Signal Connector Pinout........................................................................................................... 9
Where to Go for Support.........................................................................................................10
Software Requirements
Calibrating the FD-11603 requires the installation of NI-DAQmx on the calibration system. NI
recommends using the latest NI-DAQmx driver. Earliest driver support version for calibrating
the FD-11603 is listed in the following table.
Table 1. FD-11603 Driver Support
Driver
Earliest Version Support for Device Calibration
NI-DAQmx
18.5
You can download NI-DAQmx from
. NI-DAQmx supports many
programming languages, including LabVIEW, LabWindows
™
/CVI
™
, C/C++, C#, and Visual
Basic .NET. When you install NI-DAQmx, you only need to install support for the application
software that you intend to use.