Chapter 2
SC-2050
©
National Instruments Corporation
2-3
SC-2050 Connection
Connect your SC-2050 to the DAQ device via an SH6850 cable or
PSHR68-50 cable (for DAQCards). Connect the SC-2050 to the 100-pin
AT-MIO-16DE-10 and PCI-6071E via a 100-pin R1005050 cable,
available separately. Make this connection with the host computer powered
off. Connect one end of the ribbon cable to the SC-2050 at either connector
J1 or J2. Connect the other end to the DAQ device installed in your
computer.
Warning
Do not connect the SC-2050 to a board for which it is not designed. Such
connection can damage the SC-2050 and any or all boards/accessories connected
to the SC-2050 and host computer. National instruments is
NOT
liable for any
damages resulting from incorrect connections.
To connect your SC-2050 to the signal conditioning accessories, use a type
NB7, 26-pin ribbon cable. This cable connects the SC-2050 to any of the
SC-206X Series boards or to the 5B Series backplane. An 8-channel SSR
Series backplane with a 26-conductor ribbon cable is also available for
connection to the SC-2050. For instructions on making these connections
to the proper signal conditioning accessories, refer to
in
.
Connectors
The SC-2050 has two 50-pin ribbon cable connectors that are connected
pin-by-pin to each other. You can use either of these connectors to attach
the SC-2050 to the DAQ device via a cable. You can daisy-chain the
second connector to other 50-pin accessories.
The SC-2050 also has three 26-pin ribbon cable connectors that you can use
to connect to the signal conditioning accessories. The analog ports have pin
connections that are compatible with the 5B Series of analog signal
conditioning modules. The digital ports are compatible with the SC-206X
Series digital signal conditioners as well as with the 8-channel SSR Series
solid-state relay board.
Figure 2-2 illustrates the connections between the SC-2050 and the
necessary accessories and devices.