6703/6704 Calibration Procedure

Verifying Your Board’s Operation

Verification determines how well the board is meeting its specifications. 
By performing this procedure, you can see how your board has operated 
over time. This information can help determine the appropriate calibration 
interval for your application.

The verification procedure is divided into the major functions of the device. 
Throughout the verification process, use the tables in the 

Accuracy 

Specification

 section later in this document to see if your device needs to 

be calibrated.

Verifying Voltage Output

This procedure checks the operation of the board by verifying that the 
output of voltage channel zero is within specifications. Because of the 
channel-to-channel matching of the 6703/6704 voltage channels, if voltage 
channel zero is operating within specifications, all the voltage channels 
should be operating within specifications. However, if you want to verify 
the channels individually, this procedure can be easily expanded to verify 
all the voltage channels by performing the procedure once for each voltage 
channel. To do this, perform the procedure once for each value of the 
channel parameter from 0 to 15. Each time you repeat the procedure, you 
will also need to connect the appropriate channel to your DMM. See the

 

PCI/PXI-6703/6704 User Manual

 for the pin numbers corresponding to 

each channel. To verify the voltage output, make sure that you have read 
the 

Equipment and Other Test Requirements

 sections covered previously, 

and follow these steps:

1.

Run the 

Get_Cal_Date

 function contained in 

ni6704Cal.dll

 to 

check the board’s previous calibration date.

2.

Connect the DMM to the board as shown in the following table:

3.

Place the DMM in voltage mode.

4.

Run 

AO_VWrite

 with 

chan

 set to 0 and 

voltage

 set to 0.0. This will 

output 0.0 V on voltage channel 0.

5.

Compare the resulting value shown by the DMM to the upper and 
lower limits listed in 

Specifications

 section later in this document. 

If the value falls between these limits, the test is considered to have 
passed.

Signal Name

Description

6703/6704 Output

DMM Input

VCH0

Voltage Channel 0

Pin 34

High

AGND0

Analog Ground 0

Pin 68

Low

Summary of Contents for 6703

Page 1: ...it This section also explains internal and external calibration Equipment and Other Test Requirements This section describes what you need to do before you can calibrate your device including an over...

Page 2: ...of your application indicate how often the 6703 6704 products need to be calibrated to maintain accuracy National Instruments recommends that you perform a complete calibration at least once every ye...

Page 3: ...the calibration and verification process to approximately 10 minutes However manual calibration and verification can take as long as 1 hour You can automate the adjustment and verification procedure i...

Page 4: ...e detailed information on using the NI DAQ driver The function reference manual includes information on the functions in the driver The user manual provides instructions on installing and configuring...

Page 5: ...edure The calibration procedure in the Calibration Process section later in this document provides step by step instructions on calling the appropriate calibration functions These calibration function...

Page 6: ...as by picking up extra noise that would affect calibration Maintain the temperature between 18 28 C Keep relative humidity below 80 Use shielded copper wire for all cable connections to the device It...

Page 7: ...his procedure can be easily expanded to verify all the voltage channels by performing the procedure once for each voltage channel To do this perform the procedure once for each value of the channel pa...

Page 8: ...he procedure once for each value of the channel parameter from 16 to 31 You will also need to connect the appropriate channel to your multimeter each time you repeat the procedure See the PCI PXI 6703...

Page 9: ...make offset and gain adjustments You perform offset adjustment first followed by gain adjustment You will then perform both offset and gain adjustment again to ensure proper accuracy This second set...

Page 10: ...the voltage on voltage channel 0 Decreasing the voltage value will increase the magnitude of the voltage on voltage channel 0 You have now calibrated the voltage output Calibrating Current Output 6704...

Page 11: ...on current channel 16 Decreasing the voltage value will increase the magnitude of the current on current channel 16 You have now calibrated the current output Saving Calibration Values Once you comple...

Page 12: ...ng tables are the specifications you should use when calibrating and verifying the 6703 6704 The tables show the specifications for 1 year and 24 hour calibration intervals Using the Tables The follow...

Page 13: ...r Limit V Lower Limit V Upper Limit V 0 0 0 0003125 0 0003125 0 0003125 0 0003125 10 0 9 9992675 10 0007325 9 9990975 10 0009025 Table 2 Current Output Specifications 6704 Only Test Point Value mA 24...

Page 14: ...r additional information on the software structure Voltage Output Verification Figure 1 Voltage Output Verification Look at the 6703 6704 specification table to determine the appropriate ranges for th...

Page 15: ...tput Verification 6704 Only Look at the 6703 6704 specification table to determine the appropriate ranges for the voltages and currents to be verified Get_Cal_Date deviceNumber Connect the multimeter...

Page 16: ...IBRATION_ENABLE ND_YES AO_VWrite deviceNumber 0 0 0 AO_VWrite deviceNumber 0 10 0 6703 AO_VWrite deviceNumber 16 voltage Adjust voltage until the DMM reads 10 0 V 6704 AO_VWrite deviceNumber 32 voltag...

Page 17: ...6704 Only AO_VWrite deviceNumber 16 0 1 Connect the multimeter to current channel 16 AO_VWrite deviceNumber 33 voltage Adjust voltage until the DMM reads 0 1 mA AO_VWrite deviceNumber 35 voltage Adju...

Page 18: ...libration Procedure 18 www natinst com Saving Calibration Values Figure 5 Saving Calibration Values Set_DAQ_Device_Info deviceNumber ND_CALIBRATION_ENABLE ND_NO AO_Calibrate deviceNumber 2 1 Copy_Cal...

Reviews: