Nasco Healthcare DANi LF04500 Instruction Manual Download Page 25

Set Up

IV and Injection Arm

IV and Injection Arm 

Features included with LF04502/LF04502M, 

LF04504/LF04504M, and LF04504EX/LF04504MEX . 

DANi has skin texture realistic to touch thanks 

to modern plastics technology . The following 

adhesives will alleviate challenges in adhesion: 

3M Kind Removal Adhesive Tape, Safe N Simple 

Silicone Adhesive Tape, Curad Silicone Band-

Aid, Safe N Simple Simpurity IV Derm Silicone 

(Tegaderm equivalent) . 

Internal and External Structure

The outer skin is easily peeled off, revealing the 

“core” and veins . The skin and veins can be readily 

replaced when needed . Using smaller gauge 

needles will prolong the life of the original skin and 

veins . Replacement parts are available and listed 

at the end of this manual . The internal vascular 

structure begins at the shoulder and continues 

under the arm, crosses the antecubital fossa 

forearm, makes a loop in the back of the hand, and 

then returns to the underarm . This venous system 

is constructed of special, natural dry rubber, with 

the lumen being the approximate size of an adult 

human vein (See Figure 65) . This vascular structure 

has inlet and outlet tubing at the shoulder . It is via 

these tubes that synthetic blood is infiltrated .

Filling the Venous System and Preparing the Arm 

for Blood Draws

1 .  Prepare the synthetic blood by filling the 

pint bottle containing the synthetic blood 

concentrate with distilled water . 

2 .  Close the clamp on the IV tubing and pour the 

synthetic blood into one of the provided fluid 

supply bags . This will be IV Bag A . Fill to 500 cc 

maximum .

3 .  Hang IV Bag A no more than 18” (45 .72 cm) 

above the level of the arm . Fluid Supply stand 

shown sold separately .

4 .  Attach the tubing on IV Bag A to one of the 

shoulder tubes . Since this is a single tube loop 

system it does not matter which tube you use . 

This will now be the inlet tube .

5 .  Use the second shoulder tube for draining; this 

will be the outlet tube . With the outlet shoulder 

tube in a basin, a sink, or attached to the second 

IV Bag or IV Bag B, make sure the clamp on 

the drain tube is open . If using IV Bag B, ensure 

the clamp on IV Bag B is also open . Flush the 

vascular system with synthetic blood by slowly 

opening the clamp on IV Bag A . Allow the 

system to flush with synthetic blood until the air 

bubbles are no longer passing through the outlet 

shoulder tubing into the basin, sink, or IV  

Bag B . (See Figure 66) .

Pinch Clamp

Outlet Tubing

Inlet Tubing

Pinch 

Clamp

12"

Life/form

®

 

Blood

Venous System

Anterior

Posterior

Needle

Adapter

Posterior

Anterior

Fig 65.

Fig 66.

A

6 .  Close the clamp on the outlet shoulder tube and, 

if using IV Bag B to close off the blood outlet, the 

system is now filled and pressurized . Be sure to 

leave the clamp on IV Bag A open .

The arm is ready to practice drawing blood . 

Synthetic blood can be drawn anywhere along the 

pathway of the vein . 

25

Summary of Contents for DANi LF04500

Page 1: ...DANi Patient Care Trainer LF04500 LF04500M Patient Skills Trainer LF04502 LF04502M Patient Simulator LF04504 LF04504M Patient Simulator Export LF04504EX LF04504MEX INSTRUCTION MANUAL...

Page 2: ...my Care Lavage and Gavage 14 Male Catheterization 15 Foreskin Application Prostate Examination 16 Female Catheterization 17 Enema Administration 18 Female Cervical Exam 18 Suprapubic and Urinary Cathe...

Page 3: ...l patient for lifting and carrying Each version is affordable and durable with upgradable items DANi Teen Young Adult Patient Care Trainer General patient care daily living assistance simulation platf...

Page 4: ...uded Soft Male and Female Genitalia Normal and Cancerous Moles Stage 1 Sacral Ulcer Bandaging and Wound Dressing Recommended bandages Silicone Adhesive Bandages Patient Positioning fully articulated A...

Page 5: ...stration Left Blood Pressure Arm with Radial Pulse Injection Sites arms thigh right side buttocks supporting IM and sub cutaneous injections Gastrostomy Procedures lavage gavage tube placement Nasogas...

Page 6: ...re Urinary Catheter Care No aux equipment Stomas Standard Arms with Injection Sites Support for Oxygen Administration Left Blood Pressure Arm Option Advanced Skills Injections Sites Gastrostomy Proced...

Page 7: ...g Self Harm LF04527 Bruising Drug Use LF04526 List of Components Female Genitalia SANE LF04514 Light LF04514M Medium DANi Teen Young Adult Patient Simulator LF04504 LF04504M includes monitors DANi Tee...

Page 8: ...Hygiene Tooth brushing should be simulated without water or any cleaning agents to avoid leaking into the head of the manikin and to simplify cleanup Teeth cannot be flossed Note All Versions of DANi...

Page 9: ...ment Features included with LF04502 LF04502M LF04504 LF04504M and LF04504EX LF04504MEX Fig 6 Set Up Ear Care Bath Washing Tracheostomy Care Ear Care Both ears may be irrigated Using water to perform e...

Page 10: ...ntinue to prevent backflow into the head of the manikin Features included with LF04502 LF04502M LF04504 LF04504M and LF04504EX LF04504MEX Fig 7 Set Up Lavage Gavage Suctioning Nasogastric Carotid Puls...

Page 11: ...loop Insert the tubes back into the torso of the manikin and replace the head on the torso See Figure 14 2 Lubricate nasogastric devices and supplies being used with the Pump Spray Lubricant provided...

Page 12: ...ing a laryngoscope blade As with a real human See Figure 21 damage can be caused if improper forces and techniques are used Damaged caused by improper use will void the warranty Note There is no crico...

Page 13: ...a reservoirs See Figure 22 1 Remove the genitalia and reach through the genital cavity It is recommended to suction the fluid from the bag prior to removal to prevent spillage when removing for cleani...

Page 14: ...coupler body with bag attached onto the coupler insert part with the black O ring that is protruding from the underside of the gastrostomy inside the upper torso of the manikin You will hear a slight...

Page 15: ...Fig 30 Fig 27 Note The Fluid Supply Stand is not included Use distilled water or follow instructions on the Simulated Urine 4 Lubricate the catheter being used with the Pump Spray Lubricant provided S...

Page 16: ...te examination Following the procedure rinse the rectum with warm water and allow to dry prior to storing Note The male genital insert does not have the capacity for enema administration Must use fema...

Page 17: ...enitalia to remove any residual lubricant Fluids should be drained and the reservoir emptied after every use Ensure the reservoir is clean and dry prior to storing Female Catheterization Enema Inserti...

Page 18: ...enema reservoir holds up to 200cc of fluid Fig 41 Set Up Enema Administration Female Cervical Examination Features included with LF04502 LF04502M LF04504 LF04504M and LF04504EX LF04504MEX The female...

Page 19: ...ubricate the catheter being used with the Pump Spray Lubricant provided The catheter should be well lubricated and force should not be used when placing the tube After completion of the exercise 1 Clo...

Page 20: ...splay window are three buttons Menu Gap and Calibrate Blood Pressure Arm Features included with LF04502 LF04502M LF04504 LF04504M and LF04504EX LF04504MEX Installing the Batteries 1 Take the Blood Pre...

Page 21: ...will display in the Electronic Control Unit window See Figure 55 6 Use the up or down arrow keys located to the right of the menu button to adjust the diastolic pressure Note The systolic and diastoli...

Page 22: ...e pulse will automatically turn off when the cuff is inflated and surpasses the systolic set point It will turn on when the cuff is deflated 20 mmHg below the diastolic set point This function allows...

Page 23: ...and the main menu window will be displayed Set Up Blood Pressure Arm Fig 60 Fig 61 Performing a Blood Pressure 1 Verify the pressure line tubing from the sphygmomanometer and the audio line from the a...

Page 24: ...with the assembled sphygmomanometer 3 Obtain a new sphygmomanometer 4 Using a scissors carefully cut the tube of the sphygmomanometer about 2 from the gauge See Figure 62 5 Take the T fitting and inse...

Page 25: ...he pint bottle containing the synthetic blood concentrate with distilled water 2 Close the clamp on the IV tubing and pour the synthetic blood into one of the provided fluid supply bags This will be I...

Page 26: ...t IV needle or butterfly 7 To start the IV flow open the clamps on both IV Bags A and B Proof of proper procedure will be evidenced by the flow of fluid from IV Bag A Control flow rate with the clamp...

Page 27: ...edures and ensure clamps are open on appropriate fluid supply IV bags and or shoulder tubes 2 Look over your equipment prior to use IV tubes will kink at the clamp site with repeated use Routinely mov...

Page 28: ...aining the shoulder joint to the arm 2 Peel the skin down and remove from arm See Figure 69 3 Apply baby powder to the interior of the replacement skin and pull onto the arm 4 The key feature needs to...

Page 29: ...recautions 1 This synthetic blood is specially formulated to be compatible with the self sealing veins and plastics used in manufacturing the injectable training arm 2 NEVER use synthetic blood for in...

Page 30: ...eries into SmartScope and 2 AAA batteries into the remote control The compartments are marked as to the positions of the batteries or 4 Press the red power button on the remote control This turns on t...

Page 31: ...e speaker cord into the speaker jack on top of the SmartScope box 3 Plug the amplifier speaker into the 110V or 220V power source 4 When the speaker is connected the SmartScope earpieces will not work...

Page 32: ...ic presystolicmurmurs loud1stsound 5 Holosystolic Murmur Normal Normal Normal Normal Patienthasmitralregurgitation frequent prematureventricularcontractions Murmuriscrescendo decresendowithlate peakin...

Page 33: ...i Ronchi Ronchi Ronchi 8 Stridor Stridor Stridor StridorLowerVolume StridorLowerVolume 9 Cavernous Cavernous Cavernous Cavernous Cavernous 10 Bronchovesicular Tracheal Bronchovesicular NormalVesicular...

Page 34: ...en opening on the foot sleeve See Figure 83 Proper Application Shown See Figure 84 Set Up Pressure Injury Wounds Set Up Edema Edema Features included with LF04502 LF04502M LF04504 LF04504M and LF04504...

Page 35: ...oid imlabworld com heartisensestudent Apple Device https apps apple com us app cpr add on kitinstructor id1203164623 https apps apple com us app cpr add on kitstudent id1400115591 Powering on CPR Moni...

Page 36: ...Set Up Identifying Parts and Consumables 36 36 20 22 23 19 21 1 7 6 25 26 4 29 30 31 34 33 37 39 38 41 42 43 40 44 46 35 36 27 28 32 14 16 17 18 15 9 3 5 12 13 2 11 10 Arm Skin 24 45...

Page 37: ...t Leg Skin Medium 16 LF04518 Right Arm Skin Light LF04518M Right Arm Skin Medium 17 LF04519 Left Arm Skin Light LF04519M Left Arm Skin Medium 18 LF04521 Torso Skin Light LF04521M Torso Skin Medium 19...

Page 38: ...use Cautions Solvents or corrosive materials will damage the simulator Never place the simulator on any kind of printed paper or plastic These materials will transfer indelible stains Ball point pens...

Page 39: ...YEARS 5 Up to 5 years warranty...

Page 40: ...Nasco Healthcare 16 Simulaids Drive Saugerties NY 12477 1 833 NASCOHC 627 2642 info nascohealthcare com www nascohealthcare com PC 101038 02 23 2023 Nasco Healthcare...

Reviews: