Damage level
The frequency-dependent overload limit shown in figure 8-1 must be taken
into account in all measurements.
Fig. 8-1
Damage level
Note:
The damage level (peak) applies to a pulse duration of
≤
15.6 ms
and a duty cycle of
≤
1/64.
Measurement value collection
Sampling frequency
1.048576 MHz
Measurement interval
(measurement rate)
250 ms
(4 per second)
RMS ( RMS value detector)
Continuous integration over 4 intervals
(sliding evaluation window: 1 sec.)
Peak (Peak value detector)
Maximum value in each interval, proper phase
Measurement axes evaluated
X, Y and Z (isotropic), combined for resulting flux
density.