5 Operation and use
ELT-400
5-4
5.2
Example applications
To take advantage of the full performance of the ELT-400, it is important to
optimize the instrument settings to match the expected characteristics of the
field being investigated. The table below provides assistance for this
important task. It shows typical applications together with the appropriate
instrument settings. Please note, though, that the choice of measurement
range or the lower measurement range limit in particular must be decided
according to the actual measurement situation and the aim of the
measurement.
Application
Characteristic feature
Recommended
instrument setting
Magnetic stirring
High field strengths,
very low frequencies
•
MODE
:
Exposure STD (Exp.1/ occ.)
•
RANGE
: High
•
LOW CUT
: 1 Hz
Household
Low field strength
50/60 Hz
•
MODE
: Field strength
320
µ
T
•
RANGE
: Low
•
LOW CUT
: 30 Hz
•
DETECT
: RMS
Induction heating /
melting
High field strength
50 Hz to 20 kHz
•
MODE
:
Exposure STD (Exp.1/ occ.)
•
RANGE
: High
•
LOW CUT
: 30 Hz
Annealing / tempering
High field strength
10 kHz to 30 kHz
•
MODE
:
Exposure STD (Exp.1/ occ.)
•
RANGE
: High
•
LOW CUT
: 30 Hz
Resistance welding,
spot welding
50 Hz, pulsed)
•
MODE
:
Exposure STD (Exp.1/ occ.)
•
RANGE
: High
•
LOW CUT
: 30 Hz
MF welding
Pulsed DC field with
dominant component of
converter frequency
•
MODE
:
Exposure STD (Exp.1/ occ.)
•
RANGE
: Low
•
LOW CUT
: 1 Hz
Table 5-1
Example applications and instrument settings
Summary of Contents for ELT-400
Page 2: ......
Page 3: ...ELT 400 Exposure Level Tester BN 2304 xx Series A Operating Manual...
Page 8: ...Notes...
Page 12: ...1 Introduction ELT 400 1 4 Notes...
Page 16: ...2 Safety instructions ELT 400 2 4 Notes...
Page 22: ...3 Preparation for use ELT 400 3 6 Notes...
Page 44: ...5 Operation and use ELT 400 5 16 Notes...
Page 68: ...7 Remote control ELT 400 7 22 Notes...
Page 95: ...ELT 400 B field probe connector RS 232 Scope Charge...
Page 96: ...Notes...
Page 100: ...Notes...