ELT-400
5 Operation and use
5-11
Note:
The measurement display will fluctuate whenever the setting is
changed. This fluctuation depends on the selected lower
frequency range limit and may continue for several seconds,
particularly if 1 Hz or 10 Hz is selected. Wait until the display has
settled before starting measurements.
5.3.6
Activating the MAX HOLD display function
The default setting updates the current measurement values in the display
continuously. Activating the MAX HOLD function causes the highest level
measured since activation of the MAX HOLD function to be displayed. This
considerably simplifies the measurement of fields that vary over time, and
particularly of pulsed fields.
If required, the highest field strength existing within the volume of a room
can also be displayed by moving the instrument around the room with the
MAX HOLD display function activated.
Activating the MAX HOLD function
⇒
Press the
MAX HOLD
key until MAX appears in the display.
Deactivating the MAX HOLD function
⇒
Press the
MAX HOLD
key until MAX is no longer shown in the display.
The current measurement value is now shown in the display.
Summary of Contents for ELT-400
Page 2: ......
Page 3: ...ELT 400 Exposure Level Tester BN 2304 xx Series A Operating Manual...
Page 8: ...Notes...
Page 12: ...1 Introduction ELT 400 1 4 Notes...
Page 16: ...2 Safety instructions ELT 400 2 4 Notes...
Page 22: ...3 Preparation for use ELT 400 3 6 Notes...
Page 44: ...5 Operation and use ELT 400 5 16 Notes...
Page 68: ...7 Remote control ELT 400 7 22 Notes...
Page 95: ...ELT 400 B field probe connector RS 232 Scope Charge...
Page 96: ...Notes...
Page 100: ...Notes...