ELT-400
5 Operation and use
5-13
Possible faults:
The overload display appears
Fig. 5-6
Overload display
The selected measurement range is unsuitable for the measurement task.
1. Select HIGH range.
2. If the overload display is still shown, increase the distance of the ELT-400
from the field source.
ª
Selecting the measurement range (RANGE): see page 5-10
– or –
3. Deactivate the MAX HOLD function if it is activated.
MAX HOLD cannot be reactivated until the overload display has cleared.
ª
Activating the MAX HOLD display function: see page 5-11