background image

40

If you press YES the name inside the station will be changed.

This name it is just a label, the serial number of the area monitor will be always the same and cannot 

be changed.

If you press no, the software will ask you if you want to replace the name in the grid with that one stored 

in the area monitor.

The main window of the software show you all possible information about the area monitor and the 

probe.

Now passages to do are:

- push the button Enable Settings

- push the button Set Clock

- Hang and Exit

Summary of Contents for 8059-00

Page 1: ...AMB 8059 First Use A quick step by step guide AMB 8059 First Use A quick step by step guide...

Page 2: ...to the computer 5 The software 8059 SW02 8 Wi FI Direct AP Point to Point 14 Wi FI Remote IP FTP 20 Modem Point to Point 31 Modem FTP 37 Optical Fiber Point to Point 47 Car mounting kit 49 Area monito...

Page 3: ...fter highlighting it always with the right button of the mouse choose the voice property it should be the last one of the menu In the new window you have to choose the card called compatibility Highli...

Page 4: ...ays to choose the option Run as Administrator This is a security control introduced in Windows When you do not use this option all program files are created in a parallel folder So with the option Run...

Page 5: ...daptor This is the converter normally given with area monitor It should be automatically installed on your computer In case you can install necessary driver manually that are in the CD Use the USB con...

Page 6: ...oint to Point with the Wi Fi can be done directly only in case the area monitor and the computer are under the same Wi Fi network or through a router In case of modem connection you need a PC modem a...

Page 7: ...server and then choose between a modem connection or Wi Fi connection In case of modem connection you need a SIM card in the area monitor modem with an internet data subscription Make sure the PIN co...

Page 8: ...e COM port of your computer Press it in case you need to change the port Leave the check on the COMM and from the drop down menu choose the connection you are using with the area monitor Remember that...

Page 9: ...a note of the Terminal PASSWORD entered It will be necessary to install the control software again in case the password is missing After entering the Password with OK the software will request confirm...

Page 10: ...r the Setting Password will be requested by the software At the beginning for the first setting operations it would be helpful to leave these password blank simply pressing enter on the keyboard Set a...

Page 11: ...of the software Place the cursor on the area monitor in the grid and click once then click on button Data transfer settings The software will verify if the serial number matches or not In case you can...

Page 12: ...set to work in FTP mode and the parameters you can find in this section are of our FTP server that can be used for preliminary test Beware that our FTP server is periodically cleaned do not use it in...

Page 13: ...not mandatory In this section last 3 parameters Remote IP address Port and AP Password are used only for the over ip connection In case the Wi Fi communication is just used in FTP mode you can leave t...

Page 14: ...ame Narda_AMB_140WY60903To work in this way the area monitor hast to be set in this way Schedule Point to Point Communication Wi Fi Wi Fi Remote Address it must be 192 168 1 1 Port the port used by th...

Page 15: ...ion of your computer look for the Wi Fi network of your Area monitor and connect to it and launch it again In the first window press the button change Now check over IP and then complete the fields wi...

Page 16: ...cide to give to your monitor station As default it is the serial number of the station Insert the Telephone Number even if the connection is Wi Fi the telephone number is necessary to add tha station...

Page 17: ...onitor is directly connected to the PC you will see the following window Press YES SI Now the software will verify if the name you gave to the station is the same registered in the station s memory In...

Page 18: ...he software will ask you if you want to replace the name in the grid with that one stored in the area monitor The main window of the software show you all possible information about the area monitor a...

Page 19: ...N Example in the picture the modem will be switched ON at 13 45 and it will stay in stand by mode for 2 quarters of hours This means that for 30 minutes starting from the 13 45 the area monitor is rea...

Page 20: ...In this modality the area monitor will send all data to an FTP server where they lie up The area monitor communicates to a router thorught the Wi Fi network To work in this way the area monitor hast...

Page 21: ...AC ddress in the router s IP list The MAC address is written in the area monitor certification You have to enter in the router control panel you intend to use Look for DHCP settings and the option Add...

Page 22: ...w Insert the Station Identifier it is the name you decide to give to your monitor station As default it is the serial number of the station Insert the Telephone Number it is the phone number associate...

Page 23: ...onitor is directly connected to the PC you will see the following window Press YES SI Now the software will verify if the name you gave to the station is the same registered in the station s memory In...

Page 24: ...ame and cannot be changed If you press no the software will ask you if you want to replace the name in the grid with that one stored in the area monitor The main window of the software show you all po...

Page 25: ...of the station and it will correspond to the folder that will be created in your FTP server where you can find all data Insert the Address the IP of your FTP server Insert the User Name the user name...

Page 26: ...te with the FTP server All communications between them is monitored by the window called Driver INET FTP The first time the software will advise you that there is no folder on FTP server with that ser...

Page 27: ...our FTP server there should be a folder with the same serial number of your monitor station and inside there should be a file called creating TXT In the main window push the button Enable Settings and...

Page 28: ...data to the FTP server using the Wi Fi connection in the part Stand By you have to decide for how many quarters of hours starting from the hours of switching ON the WiFi has to try to send all data i...

Page 29: ...to this file Now we have to force the first connection between the Area Monitor and the FTP server This last operation has to be done via cable RS232 USB optical fiber using Windows HyperTerminal or a...

Page 30: ...n exhibits very long reply times which trigger Narda monitoring station TO procedures To try Narda monitoring stations and setup a first working setup we can provide you FTP space on our private serve...

Page 31: ...IM card for this way of communication need to be specifically enabled for CSD communication mode This kind of specific SIM card is normally called M2M Machine to Machine Make sure the PIN code of the...

Page 32: ...g one Now Insert the Station Identifier it is the name you decide to give to your monitor station As default it is the serial number of the station Insert the Telephone Number it is the phone number a...

Page 33: ...onitor is directly connected to the PC you will see the following window Press YES SI Now the software will verify if the name you gave to the station is the same registered in the station s memory In...

Page 34: ...the software will ask you if you want to replace the name in the grid with that one stored in the area monitor The main window of the software show you all possible information about the area monitor...

Page 35: ...hours of switching ON the modem has to stay in stand by in the part Every you have to decide the interval expressed in hours from every switching ON Example in the picture the modem will be switched...

Page 36: ...rom it the internal clock has to be set It is always better prepare the first installation on site with the help of a laptop or a tablet to verify that internal clock has the right hour and date Conne...

Page 37: ...ion Password password used for GPRS connection In this modality the area monitor will send all data to an FTP server where they lie up All you need is a PC modem a SIM card in the area monitor modem a...

Page 38: ...w Insert the Station Identifier it is the name you decide to give to your monitor station As default it is the serial number of the station Insert the Telephone Number it is the phone number associate...

Page 39: ...onitor is directly connected to the PC you will see the following window Press YES SI Now the software will verify if the name you gave to the station is the same registered in the station s memory In...

Page 40: ...ame and cannot be changed If you press no the software will ask you if you want to replace the name in the grid with that one stored in the area monitor The main window of the software show you all po...

Page 41: ...of the station and it will correspond to the folder that will be created in your FTP server where you can find all data Insert the Address the IP of your FTP server Insert the User Name the user name...

Page 42: ...te with the FTP server All communications between them is monitored by the window called Driver INET FTP The first time the software will advise you that there is no folder on FTP server with that ser...

Page 43: ...our FTP server there should be a folder with the same serial number of your monitor station and inside there should be a file called creating TXT In the main window push the button Enable Settings and...

Page 44: ...send data to the FTP server in the part Stand By you have to decide for how many quarters of hours starting from the hours of switching ON the modem has to stay in stand by and try to send all data i...

Page 45: ...PMM For further details about Device Password consult the users manual chapter 7 5 In this case with factory settings the message will be BM IDN PASSPMM BMSCGN During the connection the AMB 8059 will...

Page 46: ...n exhibits very long reply times which trigger Narda monitoring station TO procedures To try Narda monitoring stations and setup a first working setup we can provide you FTP space on our private serve...

Page 47: ...he easiest way to communicate with he area monitor Insert the USB Optical adaptor in the computer and wait for the automatic driver installation When you launch the software 8059 SW02 in the first win...

Page 48: ...the number the computer associates to the converter Now switch the main panel of the software into Point to Point and use the function direct readings to immediately receive data from the area monito...

Page 49: ...he automatic driver installation Now connect the area monitor to the computer with the 10 m optical fiber and switch it on Now launch the software NardaEMFGPSLogger exe For further explanation consult...

Page 50: ...the main window from the settings drop down menu select the option limits Edit treshold limits you need to respect and push OK In this example we have just one treshold be cause we are using a monoba...

Page 51: ...rm will warn you Now you can start to log file from the area monitor Remember that every time a new log session starts all data in the area monitor s memory will be permanently deleted In case of doub...

Page 52: ...n like probe s name GPS coordinates and field values The 8059Data gpx is the file that will be used to create the track on the map The file probe name _limits dat is just the limit associated at that...

Page 53: ...the instant value of every band Behind the field value there are general information latitude longitude and battery status On the left there is the status of GPS connection when it is green the conne...

Page 54: ...to determinate the size of the measured field on the track For example I set my treshold limit at 6 V m but the field limit is 20 V m On the track i will have a circle for every acquisition Higher is...

Page 55: ...ree website GPS visualizer http www gpsvisualizer com to convert GPX file into a map Use the browse button to select the file from your computer Leave the output format on Google Map and then click on...

Page 56: ...at radiate even for miles If the signal we are detecting on a roof is already low it will be even more inside the building or at a distance greater than the emission point Instead if the detected sign...

Page 57: ...y wet it was completely covered with water How does the refraction of the signal change It should be remembered that the area monitor measures a true value If in a point the probe marks 10 V m it is b...

Page 58: ...ands described in the user s manual chapter 4 and also verify the answer from the monitor station When I open the software some sections are not active Verify to have pushed the button Enable setting...

Page 59: ...s regarding Rate Settings and probe model When in the same day one or both these parameters change we have a not homogeneous data in that specif day The software will show you only data before the set...