background image

JuLI™Stage, Real-Time Cell History Recorder

4

5

Position Calibration 

(Vessel type)

Stitching Image Calibration

(Stitching & Map mode)

1. Press the Move To A01 or B02 button.
     Like the picture, stage will move to A01 or B02 well      
     and the  center square displays the current location.

2. Press the Check A01 or B02 button.

If you press the button, The Stage will move to 4 quad-
rants (top, bottom, left and right) to check positioning 
around the radius of the well. The outside of well will be 
displayed on each square of quad rant. If location adjust-
ment is needed, adjust is the location to be symmetrical 
based on the red line. When you double click the loca-
tion, the red line will move to the clicked location.

3. To apply the adjusted value, click Calibration A01 or B02 
    button.
    To check the current applied condition, press Check A01
    or B02 button one more time.

4. Press the Move to B02 or C03
    (Well number can be changed by vessel type. Basically, it is 
     expressed in the end well of the relevant vessel type.)

5. Press the Check B02 or C03
    Repeat the same process with A01 or B02 well calibration.

6. Press Save button and X button to finish vessel calibration. 

1. Select Make Map of Map Mode from Preview tap. 

If there is overlapped image during map mode or stitching, use this function.

2. Press Calibration button in the Make Map.

3. Select Horizontal Axis Mode button. 

4. Press Calibration distance button to adjust Y-axis 
    location. Move to appropriate location by dragging 
    the two overlapped images with the mouse.

5. Press Vertical Axis Mode button. 

6. Press Calibration distance button to adjust X-axis 
    location. Move to appropriate location by dragging 
    the two overlapped images with the mouse.

7. Press Save button and X button to finish stitching
    image calibration.

7. Click X button to end calibration.

Summary of Contents for Juli

Page 1: ...eong si Gyeonggi do 18531 Korea Tel 82 2 6220 7940 Fax 82 2 6220 7999 NanoEnTek Inc USA 240 Bear Hill Road Suite 101 Waltham MA 02451 USA Tel 1 781 472 2558 Fax 1 781 790 5649 www julistage com sales...

Page 2: ...ial foam from the box 2 Carefully lift the instrument out of the box using the handholds in the bottom of the base 3 Place the instrument on a flat level surface table or incubator 4 Unscrew the filte...

Page 3: ...ed condition press Check A01 or B02 button one more time 4 Press the Move to B02 or C03 Well number can be changed by vessel type Basically it is expressed in the end well of the relevant vessel type...

Page 4: ...first well to end well Image capture Save and overlay 8 Find optional focus adjusting the Focus bar and buttons 9 Press the Capture button to start capturing images 10 Press the Save button to save t...

Page 5: ...lease refer to the previous page Refer to the detail in the manual User manual Page 22 27 Refer to the detail options of Monitoring option tap in the Option menu User manual page 36 Operation type 1 f...

Page 6: ...tion menu User manual page 36 Operation type 2 for selecting Individual samples using the Manual focus 1 Click a well you want to capture using the Right mouse button 2 Once the well is selected stage...

Page 7: ...yeonggi do 18531 Korea NanoEnTek Inc USA 240 Bear Hill Road Suite 101 Waltham MA 02451 USA Email Website Tel 1 781 472 2558 Fax 1 781 790 5649 sales nanoentek com www julistage com www nanoentek com N...

Reviews: