ADAM-CellT, A New Standard of Automated Cell Counter
21
Introduction
Measure
Result Analysis
- Error code
Error code
Cause
E
Frames with errors are over 50% of total counting
frame.
O
Cells are more than 4X10
6
cells/mL.
H
Cells are more than 2X10
6
cells/mL.
L
Cells are less than 4X10
5
cells/mL.
U
Cells are less than 5X10
4
cells/mL.
Error frame [#]
Frame with error that contains cells whose diameter
is larger than 100μm.
When this error shown in result window, please check
the image.
Error code
Cause
E
Frames with errors are over 50% of total counting
frame.
O
Cells are more than 2X10
7
cells/mL.
H
Cells are more than 1X10
7
cells/mL.
L
Cells are less than 4X10
5
cells/mL.
U
Cells are less than 5X10
4
cells/mL.
Error frame [#]
Frame with error that contains cells whose diameter
is larger than 100μm.
When this error shown in result window, please check
the image.
[Solution T
PI
/N]
[Solution T
AO
/N]
• Please use the solution T
AO
/N when the cell concentration is above the
range of 4x10
6
cells/ml.