6
ZetaMix System
®
4.
Assembling
-
Install the furnace according to the zetasinter chamber furnace facility guide.
-
Remove the plastic cover from the heating elements under the top cover (remove the screws
to do this).
-
Install the alumina tube. Put the refractory blocks on both sides of the tube (cf. Alumina tube
chapter).
-
Close the circuit breaker, turn on the power
→
the green light comes on and the cooling fan
starts to work.
-
Turn the
“
lock
”
button clockwise
→
the temperature control unit is now switched on.
-
Press the
“
Turn On
”
button (green light is now on).
Temprature controller
Turn ON
Turn ON
Lock
Cycle
Left, Down, Up arrows
Selection screen
T/time screen
Danger due to electrical current.
During installation and maintenance
work, the electric supply to the furnace
must be switched off.
Work on the electrical equipment may
only
be
performed
by
qualified
personnel.