NANOE Zetamix ZetaSinter Operation Manual Download Page 16

 

16 

ZetaMix System

 ®

   

 

9.

 

Quick start procedure 

 
 

HEATING CURVE SETTING 

1.

 

Turn on the temperature controller by turning the LOCK button. 

2.

 

The temperature controller is initializing and then displays the initial screen. 

3.

 

Press the LETF ARROW button to display the heating curve setting. 

4.

 

The  screen  is  now  displaying  the  first  segment  (

C  01

)  and  the  associated  temperature.  This 

temperature can be modified by pressing the UP or DOWN ARROWS. After pressing the UP or 
DOWN ARROWS for a few second, the decimal point will be displaced to the left in order to 
change the temperature faster. With the side arrows, you can choose to change the number 
of units, tens, hundreds... 

5.

 

Once the first temperature is set, press the  CYCLE button to move to the first segment time 
setting  (

T  01

).  You  can  change  the  time  displayed  in  minutes  by  pressing  the  UP  or  DOWN 

ARROWS.  With  the  side  arrow,  you  can  choose  to  change  the  number  of  units,  tens, 
hundreds... 

6.

 

Then you can cycle through the segment by pressing the CYCLE button to build the desired 
heating curve. The number displayed by the letter C or T corresponds to the number of the 
segment. 

7.

 

Note that a plateau can be added by choosing the same temperature as the temperature of 
the previous segment. If so, the time setting of that segment corresponds to the time of the 
plateau. 

8.

 

Finally,  when  all  segments  are  built,  you  must  add  a  final  step  by  choosing  -121  as  the 
temperature of the last segment and not assigning any time control to this step. 

9.

 

In order to exit the heating curve setting, you can wait 20s and the temperature controller will 
automatically return to the initial screen. You can check your heating curve by entering the 
heating curve setting again (back to point 3). 

RUN A HEATING CURVE 

1.

 

To run the selected heating curve, you must first press the TURN ON button to close the circuit 
breaker. The green button will light up. 

2.

 

Press the DOWN ARROW (run) for more than one second to start the heating curve. 

3.

 

You can hold the heating curve by pressing the DOWN ARROW for a little longer than one 
second.  To  resume  the  heating  curve,  press  the  DOWN  ARROW  again  for  more  than  one 
second.  You  can  stop  the  heating  curve  by  pressing  the  UP  ARROW  for  more  than  one 
second. 

 
 
 
 
 

Summary of Contents for Zetamix ZetaSinter

Page 1: ...ry 2022 NANOE SAS C O N T A C T N A N O E C O M 3 3 9 8 1 9 8 3 3 6 4 6 R U E D E S F R E N E S 9 1 1 6 0 B A L L A I N V I L L I E R S F R A N C E W W W N A N O E C O M High Temperature chamber furna...

Page 2: ...before using Content 1 Furnace Description 3 2 Technical Specifications 3 3 Safety 4 a Intended use b Assembling c Operation d Maintenance 4 Assembling 6 5 Temperature controller 8 6 MoSi2 HeatinG el...

Page 3: ...s are to be placed inside the chamber on ceramic plates or crucibles 2 Technical Specifications Heating Power 9kW Voltage Frequency Current 3 phases neutral 380V 50Hz 32A Maximum temp 1600 C Working T...

Page 4: ...ation use and maintenance must be done according to this operation manual b Assembling Assembling must be done according to this operation manual Danger due to electrical Current During installation a...

Page 5: ...furnace temperature is above 500 C After replacing the temperature controller its settings must be adjusted before use Security warning Do not touch the inner or outer surface of the furnace during o...

Page 6: ...uit breaker turn on the power the green light comes on and the cooling fan starts to work Turn the lock button clockwise the temperature control unit is now switched on Press the Turn On button green...

Page 7: ...20 C03 300 t03 65 C04 100 t04 121 For the first high temperature heating above 400 C heating element must be passivated for 2 hours at 1200 C cf heating element chapter Using the temperature controlle...

Page 8: ...ure 50 C at room temperature SV StoP is alternately displayed on the lower window Process value Thermo couple temperature C P V Setting value temperature C Program state S V Setting key confirm KEY Da...

Page 9: ...ve to the next parameter The program parameters are displayed in the following order setpoint1 time1 setpoint2 time2 Press and hold to return to the previous setting The program step can be changed at...

Page 10: ...10 ZetaMix System Specific settings Time Setting...

Page 11: ...un Hold states Run state In Stop state press and hold until the SV window displays the run symbol The controller runs the defined program Hold state In Run state press and hold until the SV window dis...

Page 12: ...o form a protective layer against oxidation new heating elements must be heated to 1200 C for 2 hours with a ramp of 3 C min cf 4 Assembling This process should be done when using a Zetasinter for the...

Page 13: ...f the furnace chamber Then follow the procedure to tighten the screw and complete the heating element replacement 7 Troubleshooting Failure Code 1 Code 2 Explanation Solution No power The furnace is n...

Page 14: ...el The green power light is off and the red open circuit light is off Same as the previous failure and the red light is broken Same as the previous failure and change the red light The current of the...

Page 15: ...15 ZetaMix System 8 Electrical schematic diagram 380V Electrical schematic diagram...

Page 16: ...ld the desired heating curve The number displayed by the letter C or T corresponds to the number of the segment 7 Note that a plateau can be added by choosing the same temperature as the temperature o...

Page 17: ...bed above is in conformity with the relevant Union harmonisation legislation EU directive on low voltage electrical equipments 2014 35 UE EU directive on machinery 2006 42 EC 6 Applicable standards fo...

Reviews: