NANOE Zetamix ZetaSinter Operation Manual Download Page 13

 

13 

ZetaMix System

 ®

   

1)

  2)

  3)

 

 Unscrew the clip from the ceramic jaws that hold the heating element. 

 Remove the ceramic jaws and the refractory block. 

 Remove the heating element which needs to be replaces 

 Replace it with a new heating element. Replace the refractory block and ceramic jaws,  make 

sure they all fit in the same position as before.

 

4)

 

5)

 

6)

  7)

 

 

 When tightening the clip screw, make sure that the bottom of the heating element does not touch 

the bottom of the furnace chamber.   

  Then  follow  the  procedure

④③②①

to  tighten  the  screw  and  complete  the  heating  element 

replacement. 
 

7.

 

Troubleshooting 

Failure 

Code 1 

Code 2 

Explanation 

Solution 

No power 

The furnace is not 

connected to the 

power supply. 

Check the power supply 

and the electrical 

connection. 

Check and close the 32A 

circuit breaker on the 

furnace body (left). 

Lock key is 

turned 

clockwise 

(right) but the 

temperature 

controller is not 

The control circuit 

is not connected 

to the power 

supply because its 

2A circuit breaker 

is open. 

Identify (if possible) the 

cause of the failure. 

Check and close the 2A 

circuit breaker inside the 

furnace (behind the control 

panel). 

Summary of Contents for Zetamix ZetaSinter

Page 1: ...ry 2022 NANOE SAS C O N T A C T N A N O E C O M 3 3 9 8 1 9 8 3 3 6 4 6 R U E D E S F R E N E S 9 1 1 6 0 B A L L A I N V I L L I E R S F R A N C E W W W N A N O E C O M High Temperature chamber furna...

Page 2: ...before using Content 1 Furnace Description 3 2 Technical Specifications 3 3 Safety 4 a Intended use b Assembling c Operation d Maintenance 4 Assembling 6 5 Temperature controller 8 6 MoSi2 HeatinG el...

Page 3: ...s are to be placed inside the chamber on ceramic plates or crucibles 2 Technical Specifications Heating Power 9kW Voltage Frequency Current 3 phases neutral 380V 50Hz 32A Maximum temp 1600 C Working T...

Page 4: ...ation use and maintenance must be done according to this operation manual b Assembling Assembling must be done according to this operation manual Danger due to electrical Current During installation a...

Page 5: ...furnace temperature is above 500 C After replacing the temperature controller its settings must be adjusted before use Security warning Do not touch the inner or outer surface of the furnace during o...

Page 6: ...uit breaker turn on the power the green light comes on and the cooling fan starts to work Turn the lock button clockwise the temperature control unit is now switched on Press the Turn On button green...

Page 7: ...20 C03 300 t03 65 C04 100 t04 121 For the first high temperature heating above 400 C heating element must be passivated for 2 hours at 1200 C cf heating element chapter Using the temperature controlle...

Page 8: ...ure 50 C at room temperature SV StoP is alternately displayed on the lower window Process value Thermo couple temperature C P V Setting value temperature C Program state S V Setting key confirm KEY Da...

Page 9: ...ve to the next parameter The program parameters are displayed in the following order setpoint1 time1 setpoint2 time2 Press and hold to return to the previous setting The program step can be changed at...

Page 10: ...10 ZetaMix System Specific settings Time Setting...

Page 11: ...un Hold states Run state In Stop state press and hold until the SV window displays the run symbol The controller runs the defined program Hold state In Run state press and hold until the SV window dis...

Page 12: ...o form a protective layer against oxidation new heating elements must be heated to 1200 C for 2 hours with a ramp of 3 C min cf 4 Assembling This process should be done when using a Zetasinter for the...

Page 13: ...f the furnace chamber Then follow the procedure to tighten the screw and complete the heating element replacement 7 Troubleshooting Failure Code 1 Code 2 Explanation Solution No power The furnace is n...

Page 14: ...el The green power light is off and the red open circuit light is off Same as the previous failure and the red light is broken Same as the previous failure and change the red light The current of the...

Page 15: ...15 ZetaMix System 8 Electrical schematic diagram 380V Electrical schematic diagram...

Page 16: ...ld the desired heating curve The number displayed by the letter C or T corresponds to the number of the segment 7 Note that a plateau can be added by choosing the same temperature as the temperature o...

Page 17: ...bed above is in conformity with the relevant Union harmonisation legislation EU directive on low voltage electrical equipments 2014 35 UE EU directive on machinery 2006 42 EC 6 Applicable standards fo...

Reviews: