![NANOE Zetamix ZetaSinter Operation Manual Download Page 13](http://html1.mh-extra.com/html/nanoe/zetamix-zetasinter/zetamix-zetasinter_operation-manual_3210285013.webp)
13
ZetaMix System
®
1)
2)
3)
④
Unscrew the clip from the ceramic jaws that hold the heating element.
⑤
Remove the ceramic jaws and the refractory block.
⑥
Remove the heating element which needs to be replaces
⑦
Replace it with a new heating element. Replace the refractory block and ceramic jaws, make
sure they all fit in the same position as before.
4)
5)
6)
7)
⑧
When tightening the clip screw, make sure that the bottom of the heating element does not touch
the bottom of the furnace chamber.
⑨
Then follow the procedure
④③②①
to tighten the screw and complete the heating element
replacement.
7.
Troubleshooting
Failure
Code 1
Code 2
Explanation
Solution
No power
/
/
The furnace is not
connected to the
power supply.
Check the power supply
and the electrical
connection.
Check and close the 32A
circuit breaker on the
furnace body (left).
Lock key is
turned
clockwise
(right) but the
temperature
controller is not
/
/
The control circuit
is not connected
to the power
supply because its
2A circuit breaker
is open.
Identify (if possible) the
cause of the failure.
Check and close the 2A
circuit breaker inside the
furnace (behind the control
panel).