69
EN
MXA51006
(85956141BM03SJ)
Date of issue: 2022-10-10 (ver. 6)
Test
frequency
MHz
Band
a)
MHz
Service
a)
Modulation
b)
Maximum
output
W
Distance
m
Immunity
test level
V
.m
-1
385
380 to 390
TETRA 400
Pulse
modulation
b)
18 Hz
1.8
0.3
27
450
130 to 470
GMRS 460 FRS 460
FMc) ± deviation
5 kHz
1 kHz sinusoidal
waveform
2
0.3
28
710
704 to 787
LTE Band 13,17
Pulse
modulation
b
217 Hz
0.2
0.3
9
745
780
810
800 to
960
GSM 800/900
TETRA 800
iDEN 820
CDMA 1900
DECT
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
870
930
1720
1700 to
1990
GSM 1800/1900
CDMA 1900
DECT
LTE Band 1,3, 4, 25
UMTS
Pulse modulation
b)
217 Hz
2
0.3
28
1845
1970
2450
2400 to
2570
Bluetooth
WLAN
802.11 b/g/n
RFID 2450
LTE Band 7
Pulse modulation
b)
217 Hz
2
0.3
28
5240
5100 to
5800
WLAN 802.11
a/n
Pulse modulation
b)
217 Hz
0.2
0.3
9
5500
5785
NOTE: If necessary to achieve the immunity test level, the distance between the transmitting antenna and the
EM device or EM system can be reduced to 1 m. According to IEC 61000-4-3, a distance of 1 m is permitted.
a) For some services, only uplink frequencies are included.
b) The carrier wave must be modulated using a rectangular signal fi ll factor of 50 %.
c) As an alternative to FM, 50% pulse modulation at 18 Hz can be used, because even if it does not represent
real modulation, it would be the worst case.
15.3
IMMUNITY REQUIREMENTS - INPUT AND OUTPUT THROUGH THE DEVICE’S COVER FROM RF DEVICES
Summary of Contents for BM-03