False alarms - the monitor
sounds an alarm although
the child is breathing
regularly
1.
Whilst breathing the
child’s movement did not reliably reach the sensor pad.
•
For very small babies, the sensor pad can be placed directly under
the sheet (but always use an insulating pad to prevent liquids leaking
into the pad). Placing it here minimizes the likelihood of false alarms.
When the baby starts to move in the cot, place the sensor pad under
the mattress. If the baby lies in an inclined position (if it should have
its head raised as part of a doctor’s recommendation), there should
be good mechanical contact between the baby, the mattress and the
sensor pad. Prop up the cot base (not just the mattress) to meet this
condition. Or prop up the back legs of the cot.
•
Check that the mattress really lies with its entire weight upon the
sensor pad. The mattress must not be clamped tightly against the
walls of the cot so that it does not “fl oat” above the cot’s base.
2. The sensor pad does not sense the baby’s breathing properly - the sensor in
the pad is probably
mechanically damaged
(for example due to a fall) or
the
pad’s life span has expired
(2 years) - the sensor pad needs to be replaced.
There was no alarm after
taking the baby out of
the cot
The sensor pad detects disturbance in the surroundings that need to be
removed. Follow section 4.
After switching on, the
device indicates empty
batteries
Make sure you have not used rechargeable batteries (they have a lower voltage
and the device evaluates the situation as low batteries). Only alkaline batteries
can be used.
The lamp does not light
The lamp is deactivated at the factory - follow section 2.2.2. to activate it. The
lamp is also deactivated when the batteries are low. If you want to continue using
it, replace the batteries.
I placed the child on the
pad in standby mode, but
the unit does not signal
the child being placed on
the pad
This is correct behavior of the monitor. The monitor is equipped with an intelligent
evaluation of the stimulation on the pad, which constantly monitors the environment
and tries to distinguish the probable movements of the child on the pad from the
surrounding stimulation. It is also equipped with a timer, which repeats the signaling
of the laying of the baby only if no signal has been detected for a certain time
(10 secs) since the previous signaling. Therefore, the monitor will evaluate that the
child may have actually been placed on the pad, and only then will it alert you of the
possible need to switch the monitor to standby mode. Also check that you have not
deactivated the Switch Guard function (see chapter 2.2.1) - this would be signaled by
a 2x warning beep after switching on the unit during the automatic functionality test.
When the batteries are
inserted, there is no
automatic functionality test,
the monitor does not turn
on when the mechanical
button is pressed.
You have probably inserted batteries with a very low voltage state.
Replace both batteries.
The baby is in the crib and
asleep, yet the indicator
light with the power-on
symbol randomly fl ashes
orange accompanied by a
quiet beeping sound.
This is a correct function of the monitor. The monitor is equipped with a sensor
that receives and evaluates the vibrations on the sensor pad caused by breathing.
It has an implemented function of detecting the child‘s removal from the bed. If
the pad has not detected any breathing movements for 10 secs, it alerts the user
to turn off the monitor. This happens either if the baby has actually been taken out
of the crib or has not taken a breath during this time. If the baby does not breathe
again or the user does not turn off the unit, in both cases a pre-alarm and then an
alarm will be announced after 17 secs (see chapter 2.1).
11.
PROBLEMS AND HOW TO SOLVE THEM