57
Secti on 10 Cleaning and Maintenance
Cleaning
The test table insert and the test table should be kept clean for the analyzer to operate
properly.
Cauti on:
Do not autoclave the test table or test table insert.
Do not touch the white calibrati on bar.
Routi ne Cleaning of Test Table Insert
1.
Remove insert and clean thoroughly.
2.
Rinse both sides of the table insert under running water.
3.
Dry and replace the table insert.
Periodic Cleaning of Test Table When Required
1.
Remove the test table by pulling it slowly out of the analyzer. Lift the test table
insert from the test table and drain the drip tray, if necessary.
2.
Wet a cott on swab with water and carefully clean the test table (except for the
white calibrati on bar).
3.
Dry the test table thoroughly (except for the white calibrati on bar) with a soft cloth
or lint-free ti ssue.