27
● Removing the cylinder module
6-3
Replacing the Cylinder Module
1
Pull out the battery connector from the substrate.
2
Remove the clevis bolt and the trunnion pin (Fig. 24).
4
Remove the proximity switch connector which is fitted behind the substrate.
3
Retracting the cylinder rod and creating a space by flexing the knee,
slowly pull out the cylinder module from the frame. (Pull it out until the
substrate comes above the frame.)
● Installing the cylinder module
1
Insert the proximity switch connector which exists behind the substrate.
2
Retracting the cylinder rod and creating space by flexing the knee, slowly
insert the cylinder module into the frame.
4
Flexing and extending the knee, confirm that no abnormality is detected.
3
Fix the trunnion pin and the clevis bolt (Fig. 24).
Note 1: In order to prevent the clevis bolt from getting loose, apply an
appropriate amount of adhesive (equivalent to Loctite 242) to
the threads.
Note 2: Apply an appropriate amount of grease to the O-ring of the
trunnion.
Disposal of Batteries
Dispose used batteries strictly in accordance with regulations of the local
government concerned. (This product uses a lithium battery.)