2.

 

Enter calibration mode. Press [ON/OFF] to turn the power off. Press and hold [ZERO] first, then 
Press [ON/OFF] while keeping the [ZERO] button depressed, the display will show the A/D value 
(a series of random numbers).  

3.

 

Calibration by the weight. Wait for the stable indicator to be displayed, Press [SET] the display 
will show "00SAVE", after 2 seconds, the display will return to the A/D value.  

4.

 

Place the correct calibration weight on platform (see note above )., then immediately press 
[SET], the display will show "00SAVE" and calibration is finished.  

5.

 

Press [ON/OFF], turn off the power, calibration is complete. 

   

Counting procedures 

1.Press [ON/OFF] to turn on the scale.

 

Wait for "0" to appear on the display. 

2.Start the Count Procedure

 

If necessary, press 

[ZERO]

 key to set the display to “0”. 

3.

 Place a given number of samples of an item on the pan (the Sample Size should be either 10

20

50 

or 100 pieces). The weight of these samples will show on the display. 

4.

 Press the 

[SET]

 key several times to put the scale in PCS mode (the indicator should be on pcs).  

5.

 Select the sample size (the same as you chose above) by pressing the [PCS] key (press it as many times 

as necessary to put it in the correct sample size (the sample size is the same as in step three =10

20

50or100),  

6.

 Press the [SET] key, the display will show “pass”, then after 2 seconds, the scale will remember the 

sample size you selected and show the starting sample size on the display  

(you can now remove the samples if you want to return the scale to 0) 

7.

 Place the items that you want counted onto the tray and the total number of items will show on the 

display. 

8.

 Press the “SET” key to exit the counting function and return to normal weighing or you can press 

ON/OFF to turn the scale off . 

NOTE: the weight of unit sample > 10e. 

   

Summary of Contents for iBalance I11000

Page 1: ... on You can use a paperclip to slightly bend the battery prongs to make them have a better connection Also some poorly designed batteries have recessed or partially obstructed battery terminals This may cause your prongs to be touching the plastic housing of the battery instead of the metal battery terminal A Faulty AC adaptor can cause your scale to act unstable with numbers jumping all around Pl...

Page 2: ... 24 Units in one box 48 48 22 cm3 Operating Temperature 0 40c 32 104f Power source 6 AA Size Batteries or AC DC Adapter 9V 100mA optional CALIBRATION When to calibrate calibration is RARELY required Calibration may be required when the scale is first set up for use or if the scale is moved to a different altitude or gravitation This is necessary because the weight of a mass in one location is not ...

Page 3: ...a given number of samples of an item on the pan the Sample Size should be either 10 20 50 or 100 pieces The weight of these samples will show on the display 4 Press the SET key several times to put the scale in PCS mode the indicator should be on pcs 5 Select the sample size the same as you chose above by pressing the PCS key press it as many times as necessary to put it in the correct sample size...

Page 4: ...tor Indicates when the scale zero is correct Weighing unit Stable indicator Indicates when the reading is stable Low battery indicator PREPARATION Installing batteries Connecting the AC adaptor Remove the battery compartment and insert six Dry batteries R6P LR6 AA size into the battery compartment Use extreme care that the polarities plus and minus are correct Plug the AC adapter to the AC adapter...

Page 5: ...yed and press ZERO Place the objects to be weighed in the container When the reading becomes stable the stable indicator is displayed FUNCTION SETTINGS 2003 Current models only 1 Enter function setting mode Press ON OFF to turn the power off Press Mode first then press ON OFF while keeping Mode pressed the display will show b_on arrow direct to light this means enter the selection of auto backligh...

Reviews: