background image

4

5

7.  If the pH 7 calibration is successful, the display will indicate “SAVEd”, then  

“PUSHCONT” will be displayed.

8.  Push and release the push button to continue. The LED will begin flashing rapidly.

9.  Repeat steps 5 through 7 with pH 4 or 10 buffer calibration solution. 

10.  Leave  the  pen  in  the  same  buffer  solution  until  the  unit  powers  off.  The 

offset will be applied to the remaining calibration point.

11.  Verify calibration by retesting the calibration solution. 

D.  1-Point Calibration

Use pH 7, 4 or 10 buffer for 1-point calibration.

1.  Thoroughly  rinse  the  pen  by  submerging  the  sensor  in  pH  buffer  rinse 

solution and swirling it around.

2.  Push and release the push button to turn the unit on. 

3.  Push and hold the push button. The display will alternate between “CAL”, 

“FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”. 

4.  Release  the  button  when  “CAL”  displays. The  display  will  indicate  “CAL” 

and the LED will flash rapidly.

5.  While the LED flashes rapidly, dip the pen in pH buffer calibration solution 

so that the sensor is completely submerged.

6.  While the LED flashes slowly, the pH calibration point will display along with “CAL”; 

swirl the pen around to remove bubbles, keeping the sensor submerged. 

7.  If the pH calibration is successful, the display will indicate “SAVEd”, then  

“PUSHCONT”  will  be  displayed.  “PUSHCONT”  will  not  display  if  you 

calibrated 4 or 10.

8.  Leave  the  pen  in  the  same  buffer  solution  until  the  unit  powers  off.  The 

offset will be applied to the remaining calibration points.

9.  Verify calibration by retesting the calibration solution. 

E.  Factory Calibration

When pH buffers are not available, the PT2 can be returned to factory default 

calibration using the FAC CAL function. This will erase any stored wet calibration. 

NOTE: default factory calibration resets the electronics only and does NOT take 

the condition of the sensor into consideration.

To  return your unit to factory calibration: 

1.  Push and release the push button. 

2.  Push and hold the push button. The display will alternate between “CAL”, 

“FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”. 

3.  Release  the  button  when  “FAC  CAL”  displays.  The  display  will  alternate 

between “PUSHnHLD” and “FAC CAL”. 

4.  Push and hold the push button. “SAVEd FAC” displays indicating the pen 

has been reset to its factory calibration. 

MAINTENANCE

I.  Battery Replacement

The PT2 display has a battery indicator that depicts the life remaining in the battery. 

When the indicator icon is at 3 bars, the battery is full. When the indicator icon falls 

to 1 bar, replace the battery with an N type battery.

1.  In a CLEAN DRY environment unscrew the pen cap in a counter-clockwise 

motion.

2.  Slide the cap and battery housing out of the pen.

3.  Remove the depleted battery out of its housing. 

4.  Insert a new battery into the battery housing oriented with the negative end 

touching the spring.

5.  Align the groove along the battery housing with the guide bump inside the pen 

case and slide the battery housing back in.

6.  Screw the pen cap back on in a clockwise direction. Do not overtighten.

Battery Housing

Battery

+

_

Align groove in battery housing with guide 

bump in pen case.

Pen Case