II.  Routine Maintenance

1.  ALWAYS rinse the pH sensor with clean water after each use.

2.  ALWAYS replace the soaker cap half

 filled with pH Sensor Storage Solution to 

prevent the sensor from drying out after each use. 

3.  If the sensor becomes dirty, clean the sensor surface with an isopropyl soaked 

cotton swab. Then rinse thoroughly with clean water.

4.  Do not drop, throw or otherwise strike the pen. This voids the warranty.

5.  Do not store the pen in a location where the ambient temperatures exceed its 

specified Operating/Storage Temperature limits.

III.  pH Sensor Replacement 

(Follow the instructions that come with your Replacement Sensor.)

CAUTION:

 Only Remove/Replace the pH sensor in a CLEAN and DRY environment!

To remove the pH sensor:

 Remove the soaker cap; make sure the PT2 (including the 

pH sensor) is clean and dry. Loosen the battery tray 

(to allow pressure equalization). 

Then firmly grasp the pH sensor body and slowly pull the pH sensor out.

To install a new pH sensor: Line up the alignment tabs on the pH sensor with the 

alignment slots on the PT2 unit. Gently push the pH sensor into position, then 

close the battery cap.

ACCESSORIES

STANDARD SOLUTIONS FOR CALIBRATION:

The ULTRAPEN™ PT2 requires pH 4, pH 7, and pH 10 buffer solutions for wet 

calibration and pH Sensor Storage Solution for proper storage.

Order MODEL#s: PH4, PH7, PH10, and PHSS

Replacement pH Sensor (with instructions) — Order Model: RPT2

MYRON L WARRANTY

The PT2, excluding the pH sensor, has a one (1) year limited warranty.

The pH sensor has a six (6) month limited warranty. 

PT2OM 08-13

SPECIFICATIONS

 

pH

Temperature

Range:

0.00-14.00pH

0-71°C/32-160°F

Accuracy:

± 0.01pH (After Wet Calibration)

±0.1ºC/±0.1ºF

Repeatability:

± 0.01pH

±0.1ºC/±0.1ºF

Resolution:

0.01pH

0.1ºC/0.1ºF

Time to Reading Stabilization: 10-30 seconds

Power Consumption: Active Mode 37mA, Sleep Mode 2µA

Temperature Compensation: Automatic to 25ºC

Physical Dimensions: 17.15cm L x 1.59cm D or 6.75in. L x 0.625in. D 

Weight: 54g/1.9oz. (without soaker cap and lanyard) 

Case Material: Anodized Aircraft Aluminum with Protective Coating

Battery Type: N type, Alkaline, 1.5V

Calibration Solutions: pH4, pH7, pH10

Operating/Storage Temperature: 0-55ºC or 32-131ºF

Enclosure Ratings: IP67 and NEMA6

EN61236-1: 2006 – Annex A: 2008: Electrostatic discharge to case of instrument may 

cause PT2 to spontaneously power on. If this happens, the PT2 will power itself off 

within seconds.

QUICK REFERENCE INSTRUCTIONS

1.  Push and release the push button to turn the pen on.

2.  While the LED flashes rapidly, dip the pen in sample solution so that the sensor 

is completely submerged. 

3.  While LED flashes slowly, swirl pen around to remove bubbles, keeping the sensor 

submerged. Avoid contact with sides/bottom of container, if applicable. 

 

If in LIVE mode

, while the LED is flashing slowly, the display will alternate between 

the live pH and temperature readings. Note readings for your records. 

 

If in Hold mode, when the LED turns on solid, remove the pen from the solution. 

The display will alternate between the captured pH and temperature readings. 

Note readings for your records.

MYRON L COMPANY

2450 Impala Drive

Carlsbad, Ca 92010-7226

Phone: +1-760-438-2021

Customer Service E-Mail: [email protected]

Technical Support E-Mail: [email protected]

Website: www.myronl.com

MADE IN USA 

 

 

 

 

            ©2012  MYRON  L  COMPANY

Cannon Water Technology Inc.

233 Technology Way Suite 9, Rocklin, CA 95765

Voice: 1 916-315-2691, FAX: 1 916-304-0210

[email protected], https://cannonwater.com/

Reviews: