OPERATING INSTRUCTIONS

NOTE: Selecting “ESC” from any menu immediately powers the unit off without 

saving changes.

I.  Temperature Unit Selection

The PT2 allows you to select the type of units used for displaying temperature: 

˚C (Degrees Celsius) or ˚F (Degrees Fahrenheit).

To set the preference:

1.  Push and release the push button to turn the unit on. 

2.  Push  and  hold  the  button.  The  display  will  alternate  between  “CAL”,  

“FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”. 

3.  Release the button while “ºCºF TEMP” is displayed. The display will alternate 

between “PUSHnHLD” and “ºCºF TEMP”.

4.  Push and hold the button. The display will alternate between “˚C”, “˚F” and 

“ESC”. Release the button when desired unit preference displays.

5.  “SAVEd ºC” or “SAVEd ºF” will display; then the unit will power off.

II.  pH Mode Selection

The PT2 allows you to select the pH measurement mode you prefer: 

Hold

 (once stabilized, the readings are captured then displayed) or 

LIVE

 (real-time readings are displayed continuously during measurement)

To set the pH measurement mode preference:

1.  Push and release the push button to turn the unit on.

2.  Push  and  hold  the  button.  The  display  will  alternate  between  “CAL”,  “FAC 

CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”. 

3.  Release the button when “ModE SEL” is displayed. The display will alternate 

between “PUSHnHLD” and “ModE SEL”.

4.  Push  and  hold  the  push  button.  The  display  will  alternate  between  “Hold”, 

“LIVE” and “ESC”. 

5.  Release the button when desired mode displays. 

6.  “SAVEd” will display, then the unit will power off.

III.  pH Measurement

The following table explains what the LED Indicator Light signals indicate and gives 

the duration of each signal:

LED Signal

Action

Duration

Rapid Flashing Dip pen in solution.

6 sec

Slow Flashing Measurement in process.

In LIVE mode, the readings are displayed.

In LIVE mode, note measurement values.

10-30 sec in Hold mode

60 sec in LIVE mode

Solid Light

(Hold mode 

only)

Measurement is complete. 

In Hold mode, captured readings are displayed.

In Hold mode, note measurement values.

6 sec

CAUTION: 

To measure solution at the extremes of the specified temperature or pH 

range, allow the pen to equilibrate by submerging the sensor in the sample solution 

for 1 minute prior to taking a measurement.
NOTE: If you cannot dip the pen in the sample solution, pour the sample into a 

clean container. If you don’t have a container or need to test a vertical stream of 

solution, use the scoop to hold sample solution.
1.  Rinse the pen. If measuring from a container, submerge the pen and swirl it 

around in FRESH sample solution 3 times. Alternatively, 30 seconds under a 

stream or of swirling in a body of water is sufficient to prepare the sensor for 

measurement.

2.  Remove  pen  from  solution.  (Fill  the  container  one  more  time  with  FRESH 

sample solution, if applicable.) Then push and release the push button.

3.  Grasp the pen by its case with your fingers positioned between the display and 

the battery cap to avoid sample contamination. 

4.  While the LED flashes rapidly, dip the pen in FRESH sample solution so that the 

sensor is completely submerged. If you do not submerge the sensor in solution 

before the flashing slows, allow the pen to power off and retake the reading.

5.  While the LED flashes slowly, swirl the pen around to remove bubbles, keeping 

the sensor submerged. 

 

a.  In Hold mode when the LED turns on solid, remove the pen from solution. 