4
5
QUICk reFereNCe INStrUCtIoNS
1. press and release push button to turn pen on.
2. While the LED flashes rapidly, dip the pen in sample solution so
that the cell is completely submerged.
3. While LED flashes slowly, swirl pen around to remove bubbles,
keeping cell submerged. Avoid contact with sides/bottom of
container, if applicable.
4. When LED turns on solid, remove pen from solution. Display will
alternate between measurement and temperature readings.
5. Note readings for your records.
MAINteNANCe
I. Battery replacement
The PT1 display has a battery charge indicator that depicts the life left in
the battery. When the indicator icon is at 3 bars, the battery is full. When
the indicator
icon falls to 1 bar, replace the battery with an N type.
1. In a
CLeAN dry
environment unscrew the pen cap in a counter-
clockwise motion.
2. Slide the cap and battery housing out of the pen.
3. Remove the depleted battery out of its housing.
4. Insert a new battery into the battery housing oriented with the negative
end touching the spring.
5. Align the groove along the battery housing with the guide bump inside
the pen case and slide the battery housing back in.
6. Screw the pen cap back on in a clockwise direction. Do not overtighten.
II. routine Maintenance
1. Always rinse the cell and electrodes with clean water after each use.
2. If the electrodes scale or become dirty, clean the cell by submerging the
probe end in a 1:1 solution of Lime-A-Way® and water for 5 minutes.
Then rinse thoroughly with clean water.
3. Do not drop, throw or otherwise strike the pen. This voids the warranty.
4. Do not store the pen in a location where the ambient temperatures
exceed its specified Operating/Storage Temperature limits.
ACCeSSorIeS
STANDARD SOLUTIONS FOR CALIBRATION:
The ULTRAPEN™ PT1 uses the following solutions for wet calibration.
Order MODEL#s: KCL-1800, 442-3000 (2027 ppm NaCl)
MyroN L WArrANty
Excepting the battery, the Myron L PT1 Conductivity/TDS/Salinity Pen is warrantied to be free
from manufacture defect for 1 year.
Battery Housing
Battery
+
_
Align groove in battery housing with guide
bump in pen case.
Pen Case
PT1TFM 08-13