QUICk reFereNCe INStrUCtIoNS

1.  press and release push button to turn pen on.

2.  While the LED flashes rapidly, dip the pen in sample solution so 

that the cell is completely submerged. 

3.  While LED flashes slowly, swirl pen around to remove bubbles, 

keeping  cell  submerged.  Avoid  contact  with  sides/bottom  of 

container, if applicable.

4.  When LED turns on solid, remove  pen from solution. Display will  

alternate between measurement and temperature readings.

5.  Note readings for your records.

MAINteNANCe

I.  Battery replacement

 

The PT1 display has a battery charge indicator that depicts the life left in 

the battery. When the indicator icon is at 3 bars, the battery is full. When 

the indicator

 

icon falls to 1 bar, replace the battery with an N type.

1.  In  a 

CLeAN  dry

  environment  unscrew  the  pen  cap  in  a  counter-

clockwise motion.

2.  Slide the cap and battery housing out of the pen.

3.  Remove the depleted battery out of its housing. 

4.  Insert a new battery into the battery housing oriented with the negative 

end touching the spring.

5.  Align the groove along the battery housing with the guide bump inside 

the pen case and slide the battery housing back in.

6.  Screw the pen cap back on in a clockwise direction. Do not overtighten.

II.  routine Maintenance

1.  Always rinse the cell and electrodes with clean water after each use.

2.  If the electrodes scale or become dirty, clean the cell by submerging the 

probe end in a 1:1 solution of Lime-A-Way® and water for 5 minutes. 

Then rinse thoroughly with clean water.

3.  Do not drop, throw or otherwise strike the pen. This voids the warranty.

4.  Do not store the pen in a location where the ambient temperatures 

exceed its specified Operating/Storage Temperature limits.

ACCeSSorIeS

STANDARD SOLUTIONS FOR CALIBRATION:

The  ULTRAPEN™  PT1  uses  the  following  solutions  for  wet  calibration. 

Order MODEL#s: KCL-1800, 442-3000 (2027 ppm NaCl)

MyroN L WArrANty

Excepting the battery, the Myron L PT1 Conductivity/TDS/Salinity Pen is warrantied to be free 

from manufacture defect for 1 year. 

Battery Housing

Battery

+

_

Align groove in battery housing with guide 

bump in pen case.

Pen Case

PT1TFM 08-13

MyroN L CoMpANy

2450 Impala Drive

Carlsbad, Ca 92010-7226

Phone: +1-760-438-2021

Customer Service E-Mail: [email protected]

Technical Support E-Mail: [email protected]

Website: 

www.myronl.com

MAde IN U.S.A.

Reviews: