SpeCIFICAtIoNS

Measurement Range:

1 - 9999 µS or ppm (0.0010 - 9.999 ppt salinity)

Accuracy (After Wet Calibration):

± 1% of reading

Repeatability:

< 1000 µS or ppm     ± 1 Count 

≥ 1000 µS or ppm     ± 0.3% of reading

Resolution:

Conductivity and TDS: 

0.1 for 1.0 - 99.9 µS or ppm

1 for 100 - 9999 µS or ppm

Salinity:  0.0001 for 0.0010 - 0.0999 ppt

0.001 for 0.100 - 9.999 ppt

Temperature: 0.1 ºC or ºF

Time to Reading Stabilization:

10 - 20 seconds

Active Mode Power Consumption:

30 - 100 mA

Sleep Mode Power Consumption:

2 µA

Temperature Measurement Range:

0 - 71° C or 32 - 160° F

Temperature Accuracy Displayed:

± 0.1 ºC or ± 0.1 ºF

Temperature Compensation Method:

Automatic to 25ºC

Physical Dimensions:

17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D 

Weight:

55 g or 1.94 oz.

Case Material:

Anodized Aluminum with Protective Coating

Battery Type:

N type, Alkaline

Battery Voltage:

1.5 V

Calibration Solution Point:

1800 µS KCl; 3000 ppm 442™ (2027 ppm NaCl)

Operating/Storage Temperature:

0 - 55ºC or 32 - 131ºF

Water Resistance:

IP67 and NEMA 6

ORIGINAL

EDITED FOR USE

Minimum height 5 mm

 

EN61236-1: 2006 - Annex A: 2008

Electrostatic discharge to case of instrument may 

cause PT1 to spontaneously power on. In this case, 

the PT1 will power off after several seconds.

FULL operAtINg INStrUCtIoNS

I.  Solution Selection

 

The  PT1  allows  you  to  select  from  several  preprogrammed 

measurement modes. The following table lists measurement modes 

with their corresponding parameters; temperature compensation and 

TDS conversion solution models; and units of measure.

Mode

parameter

Solution Model

Units

Cond

 kCl

 

Conductivity

potassium chloride

microsiemens (µS)

tdS 

442

Total Dissolved 

Solids (TDS)

442™ Myron L Natural 

Water Standard

parts per million 

(ppm)

tdS

 NaCl

TDS

sodium chloride

ppm

SALt 

442

Salinity

442™ Myron L Natural 

Water Standard

parts per thousand 

(ppt)

SALt 

NaCl

Salinity

sodium chloride

ppt

eSC

This is the escape function. Selecting escape exits solution 

selection without saving changes and turns the PT1 off.

To select a measurement mode:

1.  Press  and  release  the  push  button.  The  LCD  will  briefly  display  the 

firmware version then the current measurement mode. If the measurement 

parameter  and  solution  type  displayed  are  correct,  proceed  to  

II. Temperature Unit Selection. If not, proceed to step 2.

2.  Immediately  push  and  hold  the  push  button.  The  display  will  scroll 

through “CAL”, “SOL 

SEL

”, “FAC 

CAL

”, “ºCºF 

T

EMP

”, and “ESC”. Release 

the button when “SOL 

SEL

” displays.

3.  While the display scrolls through “PUSHnHLD” and “SOL SEL”, push 

and hold the push button. The display will scroll through “Cond KCl”, 

“tdS 442”, “tdS NaCl”, “SALt 442”, “SALt NaCl” and “ESC”. Release 

when the desired measurement mode displays. 

4.  “SAVEd”  displays  indicating  the  measurement  mode  is  saved  in 

memory. Allow the pen to time out to turn power off.