Myron L M-6555 Manual Download Page 4

1

3

II.  temperature Unit Selection

 

The PT1 allows you to select the type of units used for temperature 

measurements. The following table lists preference options with their 

corresponding units.

Mode

Unit Preference

C

Degrees Celsius

F

Degrees Fahrenheit

eSC

This is the escape function. Selecting escape exits temperature 

unit selection without saving changes and turns the PT1 off.

To set the preference:

1.  Press and release the push button. The LCD will briefly display the 

firmware version then the current measurement mode.

2.  Immediately  push  and  hold  the  push  button.  The  display  will  scroll 

through “CAL”, “SOL 

SEL

”, “FAC 

CAL

”, “ºCºF 

T

EMP

”, and “ESC”. Release 

the button when “ºCºF 

T

EMP

” displays.

3.  While the display scrolls through “PUSHn

HLD

” and “ºCºF 

T

EMP

”, push 

and hold the push button. The display will scroll through “C”, “F” and 

“ESC”. Release when the desired unit option displays.

4.  “SAVEd” displays indicating the unit preference is saved in memory. 

Allow the pen to time out to turn power off.

III. Normal operation

 

Before you take a reading, make sure the pen is clean, calibrated (see 

IV. Calibration) and in the appropriate measurement mode. The sample 

solution must also be within the specified measurement range. Keep 

all foreign material away from the sample to avoid contamination.

 

Note:

  If  you  cannot  dip  the  pen  in  the  sample  solution,  pour  the 

sample into a clean container. If you don’t have a sample container 

and need to test a vertical stream of solution, use the scoop. 
The following table explains what the LED Indicator Light signals mean 

and gives the duration of each signal.

Led Indicator Light Signal

Meaning

duration

Rapid Flashing

Dip pen in solution

6 sec

Slow Flashing

Measurement in process

10-20 sec

Solid Light

Note measurement value

6 sec

CAUtIoN:

  To  measure  solution  at  the  extremes  of  the  specified 

temperature range, allow the pen to equilibrate by submerging the cell in 

the sample solution for 1 minute prior to taking a measurement.

1.  Rinse the pen 3 times by submerging the cell in fresh sample solution 

and swirling it around.

2.  Remove pen from solution, then press and release the push button. 

Firmware version will be displayed, then current measurement mode. 

3.  Grasp the pen by its case with your fingers positioned between the 

display and the pen cap to avoid sample contamination. 

4.  While the LED flashes rapidly, dip the pen in fresh sample solution so 

that the cell is completely submerged. If you do not submerge the cell 

in solution before the flashing slows, allow the pen to power off and 

retake the reading.

5.  While  the  LED  flashes  slowly,  swirl  the  pen  around  to  remove 

bubbles, keeping the cell submerged. Keep the pen at least 1 inch  

(2½ cm) away from sides/bottom of container, if applicable.

6.  When the LED turns on solid, remove the pen from solution. The display 

will  alternate  between  the  measurement  and  temperature  readings. 

Note the readings for your records.