IV. Calibration

The PT1 is designed to be very reliable and requires only infrequent 

calibration.  The  Myron  L  Company  recommends  calibrating  each 

measurement  mode  you  use  once  monthly.  However,  you  should 

check the calibration whenever measurements are not as expected.

The  PT1  is  programmed  for  2  calibration  options:  Wet  Calibration 

or  Factory  Calibration.  Wet  calibration  is  most  accurate.  But  if  a 

high quality standard KCl-1800 µS or 442-3000 ppm solution is not 

available, the PT1 can be returned to factory settings.

A.  Wet Calibration

Use calibration solution specified for measurement mode: Use KCL-

1800  for  Cond 

KCl

;  Use  442-3000  for  tdS 

442

,  SALt 

442

,  tdS 

NaCl

and  SALt 

NaCl.

  See  Specifications  table  for  442  solution  ppm  NaCl 

equivalent  value.  Calibrating  tdS  simultaneously  calibrates  SALt  for 

the same value and vice versa.

1.  Pour Myron L Company calibration solution into a clean container.

2.  Rinse  the  pen  3  times  by  submerging  the  cell  in  fresh  calibration 

solution and swirling it around.

3.  Remove pen from solution, then fill the container one more time.

4.  Press and release the push button. The LCD will briefly display the 

firmware  version  then  the  current  measurement  mode.  Ensure  the 

PT1 is in the correct solution mode.

5.  Immediately  push  and  hold  the  push  button.  The  display  will  scroll 

through “CAL”, “SOL 

SEL

”, “FAC 

CAL

”, “ºCºF 

T

EMP

”, and “ESC”. Release 

the button when “CAL” displays.

6.  Grasp the pen by its case with your fingers positioned between the 

display and the pen cap to avoid sample contamination. 

7.  While the LED flashes rapidly, dip the pen in calibration solution so that the 

cell is completely submerged. If you do not submerge the cell in solution 

before the flashing slows, allow the pen to power off and start over.

8.  While  the  LED  flashes  slowly,  swirl  the  pen  around  to  remove 

bubbles,  keeping  the  cell  submerged.  Keep  pen  at  least  1  inch  

(2½ cm) away from sides/bottom of container. 

9.  When the LED light stays on solid, remove the pen from the solution.  

“CAL SAVED” will display indicating a successful calibration.

 

Note: 

If an incorrect solution is used or the measurement is NOT 

within calibration limits for any other reason, “Error” displays alternately 

with “CLEAn 

CEL

/CHEC 

SOL”

. Check to make sure you are using the 

correct calibration solution. If the solution is correct, clean the cell by 

submerging the cell in a 1:1 solution of Lime-A-Way® and water for 5 

minutes. Rinse the cell and start over.

10. Small bubbles trapped in the cell can give a false calibration. Measure 

the calibration solution again to verify correct calibration. If the reading 

is not within ±1% of the calibration solution value, repeat calibration.

B.  Factory Calibration 

 

If you do not have the proper calibration solution or wish to restore the 

pen to its original factory settings for any other reason, use the FAC 

CAL

 function to calibrate the PT1.

1.  Press and release the push button. The LCD will briefly display the 

firmware version then the current measurement mode.

2.  Immediately  push  and  hold  the  push  button.  The  display  will  scroll 

through “CAL”, “SOL 

SEL

”, “FAC 

CAL

”, “ºCºF 

T

EMP

”, and “ESC”. Release 

the button when “FAC 

CAL

” displays.

3.  While the display scrolls through “PUSHn

HLD

” and “FAC 

CAL

”, push 

and hold the push button until the display scrolls through “SAVEd” and 

“FAC 

CAL

”, indicating the pen has been reset to its factory calibration. 

4.  Allow the pen to time out to turn power off.