77
Trigger timing and
duration of
electrical
stimulation
(for Walk Mode
only)
When the patient's extension/flexion rises to -30°, the device can trigger
electrical stimulation, and when the patient's extension/flexion drops to -5°, the
device's electrical stimulation output is turned off. And the duration of electrical
stimulation is 1s-3s.
Timer:
a) Treatment time: 1-60min, increments of 1min;
b) Timer tolerance: ±2%;
c) When finished, the device can stop output and prompt.
Output Intensity:
Adjustable, 0-100mA, increments of 1mA (at 500Ω load)
Maximum Output
Amplitudes (V):
55 V at 500 Ω Load
Maximum Current
(r.m.s):
50mA r.m.s.at 500 Ω Load
2. Other Specifications
Product Name
MyndStep Foot Drop Stimulator
Model
800102