21
EMC Guidance
MyndStep needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided. This device has been thoroughly tested and inspected to
assure proper performance and operation.
⚫
Keep MyndStep away from active HF surgical equipment, and the RF shielded room of a medical
device for magnetic resonance imaging, where the intensity of EM disturbances is high.
⚫
Do not place
MyndStep adjacent to or stacked with other equipment, or it could result in improper
operation. If such use is necessary, all the devices should be observed to verify that they are
operating normally.
⚫
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) other than the smart phone to be used with MyndStep should be used no closer
than 30 cm (12 inches) to any part of MyndStep, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
⚫
Use of accessories, transducers and cables other than those specified or provided by the