Issue
Possible causes
Solutions
3. The status indicator on
the unit flashed white, and
the unit switched off.
The battery of the unit is empty.
Charge the unit (see chapter
Preparing for use).
4. There is no output.
1. Did not press the start button;
2. Did not adjust the output
intensity;
3. The electrode lead wire is
poorly connected;
4. The electrode lead wire has
been damaged.
1. Press the start button;
2. Adjust the output intensity;
3. Re-connect the electrode lead
wire;
4. Replace a new electrode lead
wire.
5. Stimulation is
uncomfortable.
1. The output intensity is too
high;
2. Electrodes are too close
together;
3. Damaged or worn electrodes
or electrode lead wires;
4. Electrode active area size is
too small.
1. Decrease the output intensity;
2. Re-position the electrodes;
3. Replace the lead wires;
4. Replace electrodes with ones
that have an active area no less
than 25.0 cm