Meter activation and battery change
1. Turn the meter over. Press and
push battery cover to open.
PRECAUTION
-
Please follow local
and discard used battery properly.
regulations
4. The meter performs a self test.
5. Press any button to exit the self test and enter setting mode.
6. Set the time and date
the batteries are replaced. See chapter "Setting up your meter
–
Test results are still
in the memory.
after
Setting the date, time and volume".
retained
3. Slide the battery cover
back until it snaps into
place.
2. Install the battery. Be sure to
put battery in correct way with
the "+" symbol up.
Meter activation and battery change
Your meter comes with two CR2032, 3 volt, batteries installed. Two new batteries will provide
power to perform about 1,000 tests under normal use. Press the main button or insert a test strip
to activate your meter.
NOTE
-
When the meter backlight
function, please change your backlight battery.
- When the backlight battery runs out of power
still work
but without
backlight
ceases to
the meter will
properly,
illumination.