47
46
TM
®
Limitations of mylife Pura meter
NOTE
- Do not use this meter close to strong electromagnetic radiation sources, in order to avoid
interference with normal operation.
- Keep the meter free of dust, water or any liquid.
Meter data transmission management (optional)
- The BGMS is not intended for serum or plasma testing (for capillary whole blood testing only).
- Inaccurate test results may be obtained at altitude of more than about 3,048 meters (10,000
feet).
- Severe dehydration may cause inaccurately low results.
- The BGMS has not been validated for use on neonates.
- The following compounds may interfere with the glucose measurement at the concentration
listed:
Ascorbic acid
≧
0.34 mmol/L
Glutathione reduced
≧
2.28 mmol/L
Uric Acid
≧
0.96 mmol/L
TM
®
The meter data can be transmitted to the computer with mylife Pura PC-Link adapter. The
®
®
data must be managed by four different softwares: Diabass , diasend , SiDiary and mylife™
Software.