mylife Pura test strip
TM
®
mylife Pura test strip
View window
This window is yellow before
applying blood sample. It gradually
turns red
with blood.
as it fills
Hand bar
Grip here to insert
test strip into meter.
Sample entry
µL
Apply a drop of blood or control solution here.
The test
only 0.75
of blood.
requires
Electrode
Sensing signal output terminals.
contacts
Gold electrodes
Electrochemical sensor
Indication symbol
Insert test strip with indication
symbol upfront and down towards
meter.
TM
®
TM
®
The meter must only be used with mylife Pura test strips and mylife Control / Pura control
solution.
The use of other test strips or control solutions may lead to incorrect results.
PRECAUTION
-
Recap the test strip vial immediately after taking out a test strip.
-
Do not reuse test
-
Do not use expired test strips.
-
Record the date of opening a new test strip vial for the first time. Discard the vial of test
strips 6 months
-
Store the test strips in a cool (4-30
℃
and dry location (<90% relative
humidity). Do not expose
-
For detailed information, please refer to the
test strips package insert.
strips. Test strips are intended to be used only once.
after first opening.
resp. 39-86°F)
to heat or direct sunlight.
TM
®
mylife Pura
- If the meter and test strips are exposed to considerable temperature variation, please
wait 30 minutes before measurement.