Skin Type I
Skin Type II
Skin Type III
Skin Type IV
Skin Type V
Skin Type VI
40ms
80ms
200ms
200ms
300ms
400ms
black
10Hz 4-6
J/ cm
2
5Hz
8-12
J/ cm
2
2Hz 26-30
J/ cm
2
1Hz 28-32
J/ cm
2
1Hz 40-44
J/ cm
2
1Hz 30-34
J/ cm
2
dark brown
10Hz 4-6
J/ cm
2
5Hz
8-12
J/ cm
2
2Hz 28-32
J/ cm
2
1Hz 30-34
J/ cm
2
1Hz 42-46
J/ cm
2
1Hz 32-36
J/ cm
2
light brown
10Hz 5-7
J/ cm
2
5Hz 10-14
J/ cm
2
2Hz 30-34
J/ cm
2
1Hz 32-36
J/ cm
2
1Hz 44-48
J/ cm
2
1Hz 34-38
J/ cm
2
blonde/red
10Hz 6-8
J/ cm
2
5Hz 12-16
J/ cm
2
2Hz 32-36
J/ cm
2
1Hz 34-38
J/ cm
2
1Hz 46-50
J/ cm
2
1Hz 36-40
J/ cm
2
3.3.2.2 Skin type setting
The ratios of light absorption for different skin types are different. ( dark skin easily absorb
light energy, light skin absorb less light energy).Weak laser power is suitable for dark skin
hair removal; strong laser power is suitable for light skin hair removal.
“
FIBER-808 I
”
concludes depilation treatment proposals for six different skin types.
3.3.2.3 Energy density (j/cm
2
)
Energy density (j/cm
2
) is the output energy per unit area. Energy density is adjustable within
certain ranges(1-80j/cm
2
is adjustable for spot size of 5*15mm). Fast mode, A,B,C,D,E have
preset energy density suitable for hair removal of different skin parts. Users can fine adjust
energy density by touching "+", or "-". Press "Standby" and "Ready" key to start the
treatment.
3.3.2.4 Frequency setting
Make setting for number of continuous pulses within one second. Frequency adjustable
ranges are different for different working modes.
3.3.2.5 Cooling system
The cooling system is used to cool the probe on the hand piece and protect it from skin burn
by overheating. Do turn on cooling system and cool the probe to a certain temperature
before operation.
After powering on machine, it is required to start cooling system for 3 to 5 minutes before
operation, Otherwise, the system would give alarm probe overheating and lock the Ready
button to cool the probe surface until it reaches the safe temperature. And then automatic
alarm disappears, unlock
“
ready
”
key.
The cooling system could be started both manually and automatically.
Automatic mode: The cooling system would start working automatically when pressing
“
ready
”
key.