background image

TROUBLE SHOOTING:

.

1. RESTRICTED OR NO WATER FLOW:

a. Ensure Water Supply service stop valve is fully open.
b. Verify minimum 20 psig supply line flow pressure.
c. Check for twist or kinks in spout tubing.
d. Check the water inlet “Y’’ strainer. Sediment from the main supply can get trapped in the
   screen along with installation materials such as pipe dope and flux.  The screen should be 
   cleaned and checked on a regular basis and replace if needed.
e. Flow control in solenoid valve outlet elbow clogged remove & clean.

SENSOR TROUBLE SHOOTING:
1. IF LIGHT WITHIN SENSOR DOES NOT FLASH ONCE WHEN USER IS WITHIN RANGE:

a. Verify 9VDC output of battery pack.
b. Replace defective batteries.
c. Sensor in “Security Mode” after 30 seconds of constant detection. Remove source of
   detection and wait 30 second before checking.
d. Sensor is picking up a highly reflective surface. Eliminate cause of reflection and wait 30
   seconds before checking.

2. IF LIGHT WITHIN SENSOR LENS FLASHES ONCE WHEN THE USER IS WITHIN RANGE:

a. Repair bad connection from sensor to solenoid.
b. There is debris or scale in the solenoid assembly. Remove solenoid, pull out plunger and
   spring. Clean with scale remover solution.
c. There is debris or scale in center or two holes in convolution of the water diaphragm. 
   Remove and clean.

CLEANING & MAINTENANCE GUIDE:
For Powder coated units:

 Units should be cleaned using a mild soap solution with a sponge or 

cotton cloth. Wipe down surfaces then rinse with clean water.

For Stainless steel units:

1. To Remove water spots or rust spots, stainless steel cleaner/polish on a cloth is
   recommended.
2. If there are stubborn spots or if you wish to treat a scratch, synthetic abrasive general
   purpose pads   such as   scotch brite are recommended.
3. Apply stainless steel cleaner/ polish to the synthetic abrasive pads and carefully rub the 
   panel with the grain.
4. DO NOT use harsh chemicals, abrasive or petroleum based cleaners. Use of these will
   void the Acorn warranty. DO NOT use abrasives on powder coated units.
5. Stainless steel should be kept clean at all times. If a coating of stainless steel cleaner/ 
   polish is maintained, stainless steel surfaces will retain their new, clean, polished 
   appearance indefinitely.
6. Periodically remove access panel of cooler and clean out inline “Y’’ strainer

MURDOCK  MFG.  • 

15125  Proctor  Avenue  •  City  of  Industry,  CA  91746  USA 

Phone  800-453-7465  or  626-333-2543  •  Fax  626-855-4860  •  www.murdockmfg.com

Member of

7013-037-001

Revised: 07/26/18

Page 13 of 16

 

I N S TA L L AT I O N / M A I N T E N A N C E  I N S T R U C T I O N S

Semi-Recessed Bottle Fillers

 

Summary of Contents for BF38 Series

Page 1: ...on parts only This warranty does not cover installation or labor charges and does not apply to materials which have been damaged by other causes such as mishandling or improper care or abnormal use The repair or replacement of the defective materials shall constitute the sole remedy of the Buyer and the sole remedy of Murdock under this warranty Murdock shall not be liable under any circumstances ...

Page 2: ...LEFT AND REAR WASTE ORIENTATION ARE OPTIONAL IF NEEDED FOR CALLOUTS INDICATED A C C A IN OUT L C L C D E F 31 2 90 1 4 PUNCHING 22 PLACES FOR ANCHORING HARDWARE BY OTHERS 3 8 OD COPPER SUPPLY INLET 1 1 4 O D WASTE OUTLET ELECTRICAL SERVICE ROUGH 3 8 OD COPPER SUPPLY INLET CHILLER 3 8 OD COPPER SUPPLY OUTLET CHILLER MF10 MOUNTING FRAME ASSEMBLY A B C D E F G 511 2 1310 A F F G Finished Floor MURDOC...

Page 3: ...FT AND REAR WASTE ORIENTATION ARE OPTIONAL IF NEEDED FOR CALLOUTS INDICATED 1 4 PUNCHING 22 PLACES FOR ANCHORING HARDWARE BY OTHERS 3 8 OD COPPER SUPPLY INLET 1 1 4 O D WASTE OUTLET ELECTRICAL SERVICE ROUGH 3 8 OD COPPER SUPPLY INLET CHILLER 3 8 OD COPPER SUPPLY OUTLET CHILLER D A C C D A ALT IN OUT L C L C D E F 48 1220 A F F FINISHED FLOOR MF10 MOUNTING FRAME ASSEMBLY A B C D E F G G MURDOCK MFG...

Page 4: ...sing 10 32 x 1 2 Hex Btn Screws provided mount bottle filler mounting frame to the MF10 frame assembly 3 Place the Chiller Unit onto the basepan of the MF10 Frame Assembly 7 Make up all power connections to the ground neutral hot within the electrical box and chiller unit 1 Insert MF10 Frame assembly into the rough in block out and secure using 1 4 20 UNC mounting hardware Provided by others 5 On ...

Page 5: ... be filled just below the filler tube in the center of the unit then push pushbutton When the container is almost filled release the pushbutton WATER TUBING CONNECTIONS NOTE Insulation on all water tubes not shown all PE water tubes are insulated MURDOCK MFG 15125 Proctor Avenue City of Industry CA 91746 USA Phone 800 453 7465 or 626 333 2543 Fax 626 855 4860 www murdockmfg com Member of 7013 037 ...

Page 6: ...NG DETAIL TO CHILLER IN INSTALLED FILTER DETAIL NOTES 1 CUT AND DEBURR WATER LINE TUBING AS REQUIRED 2 ENSURE TUBING END IS NOT DEFORMED BEFORE INSERTING IN Y STRAINER 3 CUT INSULATION AS REQUIRED TO COVER WATER TUBING REMOVE TUBING FROM Y STRAINER OUTLET INSTALL TO OUTLET OF CHILLER OUT IN MOUNT FILTER BRACKET TO CORNER OF MOUNTING FRAME FROM SUPPLY INLET 2 1 3 5 4 ITEM PART NUMBER DESCRIPTION 1 ...

Page 7: ... the inlet valve provided by others Turn valve knob 1 4 turn clockwise to close 4 Turn inlet 1 4 turn knob 1 4 turn counterclockwise to open and test for leaks and proper operation before re mounting panel assembly to fixture 2 With the filter assembly secured on the unit hold the cap firmly turn the replaceable cartridge filter counterclockwise to remove I N S TA L L AT I O N M A I N T E N A N C ...

Page 8: ...nsor wires to the solenoid valve wires See Below FOR BAT BATTERY OPERATION OPTION TRANSFORMER WIRING AND WATER TUBING CONNECTIONS MURDOCK MFG 15125 Proctor Avenue City of Industry CA 91746 USA Phone 800 453 7465 or 626 333 2543 Fax 626 855 4860 www murdockmfg com Member of 7013 037 001 Revised 07 26 18 Page 8 of 16 I N S TA L L AT I O N M A I N T E N A N C E I N S T R U C T I O N S Semi Recessed B...

Page 9: ...replaced BOTTLE COUNTER ADJUSTING RESETTING INSTRUCTIONS Bottle Counting Function The software applies a flow volume of approximately 16 9 fl oz volume in standard size plastic water bottle to each bottle counted If the flow volume is less than 16 9 fl oz there will be no count but the volume will accumulate so that part way through the next cycle the total bottle count will change Description Ill...

Page 10: ... as a filter Replace old filter with new filter Depress the Reset Mode selection button for 2 SECONDS The REPLACE FILTER will no longer appear on the display Test by actuating the bottle filler alert will not appear Systems With or Without Filter Depress the Reset Mode selection button for 6 Seconds the number of seconds will count up on the display At the end of the 6 seconds FLTR YES or FLTR NO ...

Page 11: ...7013 037 001 Revised 07 26 18 Page 11 of 16 SOLENOID CONNECTION POWER SUPPLY CONNECTION BLACK CABLES SENSOR CONNECTION BLACK RED CABLES SENSOR CONNECTION TRANSFORMER SOLENOID VALVE BOTTLE COUNT DISPLAY SENSOR BOTTLE FILLER BOTTLE COUNTER CONNECTIONS I N S TA L L AT I O N M A I N T E N A N C E I N S T R U C T I O N S Semi Recessed Bottle Fillers ...

Page 12: ...ull on the PE tubing to remove from the push in fitting TO CUT TUBE Cut to fit length of 1 4 PE tubing and remove any burrs or sharp edges Ensure that the outside diameter is free from score marks Tube ends should be square INSERTING THE TUBE 1 Firmly and fully insert the tubing end into the push in fitting up to the tube stop located approximately deep COLLET COLLET O RING O RING TUBE STOP COLLET...

Page 13: ...e is debris or scale in center or two holes in convolution of the water diaphragm Remove and clean CLEANING MAINTENANCE GUIDE For Powder coated units Units should be cleaned using a mild soap solution with a sponge or cotton cloth Wipe down surfaces then rinse with clean water For Stainless steel units 1 To Remove water spots or rust spots stainless steel cleaner polish on a cloth is recommended 2...

Page 14: ...Place the battery pack back to mounted location and test for proper operation Then close bottler filler door and lock WIRING DIAGRAM BATTERY PACK SOLENOID SENSOR RED WIRES MURDOCK MFG 15125 Proctor Avenue City of Industry CA 91746 USA Phone 800 453 7465 or 626 333 2543 Fax 626 855 4860 www murdockmfg com Member of 7013 037 001 Revised 07 26 18 Page 14 of 16 I N S TA L L AT I O N M A I N T E N A N ...

Page 15: ... O D BLUE TUBING 8 7013 102 002 DRAIN TRAY 21 7000 021 001 Y STRAINER ASSEMBLY 9 7000 006 000 DRAIN ADAPTER GASKET 22 2566 150 001 HEMISPHERICAL PUSHBUTTON 10 7003 186 199 1 1 4 x 2 3 4 PLASTIC DRAIN TUBE 23 7013 103 199 MESH SCREEN 11 4970 265 000 1 1 4 P TRAP 24 7013 114 199 SCREEN RING 12 7008 010 001 8GPH CHILLER ASSEMBLY 25 0302 005 000 1 4 20 UNC STN STL HEX NUT 13 7013 205 001 MF10 FRAME AS...

Page 16: ...C DRAIN TUBE 24 7013 019 199 FOAMTAPE 10 4970 265 000 1 1 4 P TRAP 25 7013 010 199 NANO SENSOR BRACKET 11 7008 010 001 8GPH CHILLER ASSEMBLY 26 0302 003 000 8 32 HEX NUT 12 7013 205 001 MF10 FRAME ASSEMBLY 27 7013 103 199 MESH SCREEN 13 7013 201 199 VENTED PANEL 28 7013 114 199 SCREEN RING 14 7013 128 199 10 32 x 1 2 S S HEX BTN HDSCREW 29 0302 005 000 1 4 20 UNC STN STL HEX NUT 15 0116 003 000 10...

Reviews: