Table 7.1 Humidity menu
7.1.1 Internal controlling sensor
The internal humidity sensor is the controlling sensor as default. This is indicated by a (*) before the reading
in the display (*RH1… / *D1… / *X1…).
7.1.2 External controlling sensors
Optionally, up to two external sensors can be connected. They will automatically be designated
(RH2 / D2 / X2) and (RH3 / D3 / X3) by the system (number two being the first connected in the chain).
When connected, it is possible to choose one of the external sensors as control. The display will then show
(*RH/D/X2…) or (*RH/D/X3…).
With an external sensor set as control sensor, it is possible to disconnect and reconnect the external
sensor/-s without losing the control setting, e.g. [*RH2…] while the unit is switched off (but powered).
When switched on, the system makes five attempts (during approx. 20 seconds) to contact the external
controlling sensor. If this fails, the system reverts to the internal RH1-sensor. Subsequently any renewed
assignment of the external sensor as control must be carried out manually.