10
Installation
This section address installation and bring-up of the instrument, addressing the following topics:
- System start-up
- Connection guide
System Start-up
To use the
DSO
under Windows XP, Windows 7 and Vista, it is important that the correct start-up
sequence is followed:
▪
▪
Install the DSO GUI software
▪
Connect the power cable to the power jack and field it with a +5 V/ 1.5 A supply
▪
Note power plug is already included in the package accessories.
▪
Communication through USB to the instrument is required to configure its own IP
only
▪
Install the USB driver to the PC (Appendix I)
▪
Change the IP of the DSO to fit in the network range by and using the software DSO Conf
(Appendix III)
▪
Communication through Ethernet port is required for data acquisition
▪
Connect the PC to the DSO through an Ethernet cable, no driver required
▪
Now the DSO is powered up, having the right IP, the Ethernet cable links the instrument to
the PC with the GUI correctly installed
▪
To open the GUI, double click on the software icon located in the Desktop directory
▪
In order to characterize the signals fielding the instrument, check the Data and Connection
configuration
NOTE:
This release supports Ethernet connection for the software applications, and the USB port
is used only to configure the device IP Address (note that the IP should be in the same range of the