W2100-HS8 Headstage
Pin Layout of the 8-Channel Headstage for the Wireless2100-System
multi
channel
systems
*
W2100-HS8 top side:
Please use the connector in the lower
right (storage battery) for orientation
of the headstage.
W2100-HS8 bottom side:
Connector for the electrode probe
or the ME/W-Signal generator.
Connector for W2100-HS8
with Omnetics socket A79039-001
This Omnetics connector mates
with Omnetics connectors such as:
Through-Hole:
A79038-001 (NPD-18-DD-GS)
Horizontal Surface Mount:
A79040-001 (NPD-18-AA-GS)
Vertical Surface Mount:
A79042-001 (NPD-18-VV-GS)
Cable (18.0" 34 AWG lead-wire):
A79044-001 (NPD-18-WD-18.0-C-GS)
W2100-HS8 with Omnetics connector A79039-001
(
NSD-18-DD-GS
, female 2 guide posts)
Diagram of the bottom side with pin layout.
Pin Layout
of the
Omnetics socket A79039-001
Please orientate the headstage as shown in the diagram.
Guide
post
NC
Not
connected
GND
Ground
REF
Reference
E1 to E8
Electrode 1 to Electrode 8 = MC_Rack Channel 1 to Channel 8
Storage battery
connector on the
opposite side
for orientation.
REF
GND
NC NC NC NC NC NC NC NC
Please orientate the headstage
as shown in the diagram.
REF E1 E2 E3 E4 E5 E6 E7 E8 GND