iv) Select whether the target muscle plugged into Ch 1 is on the right or left side
of the patient’s body. This appropriately adjusts the visual display on the next
screen.
v) Launch test and follow prompts to relax and contract maximally for three sets
of 5 seconds on and 5 seconds off.
vi) Select goal based on Average MVC results. Goal can be set to 100% or 75%
of MVC achieved during setup, or a custom goal can be typed in.
vii) Save to return to Settings; results of setup will be reflected in Ch 1 goal
settings automatically.
b) Channel Mode Select: Select dual or single channel as desired.
i) If in single channel mode, ensure you are plugged into Channel 1 only;
Channel 2 will not display data in single channel mode.
ii) Single channel mode is ideal for isometric exercise and basic activation
training.
iii) Dual channel mode is ideal for bilateral comparisons, A/B activation ratio
monitoring, agonist/antagonist monitoring, or compensation monitoring.
c) Match Goals Select: Toggle Match Goals ON to lock goals together for
simultaneous adjustment. Use for comparison of muscles with similar MVC levels
or goals. Toggle Match Goals OFF to adjust Ch 1 and Ch 2 goals independently.
Use for comparison of muscles with different MVC levels. Ch 1 goal is default
when Match Goals is ON.