Note:
Ensure that your involved muscle is plugged into channel 1 during the NMDT.
Electrode placement should be as identical as possible on both healthy and
involved sides of the body. You will be prompted to ensure that Match Goals is ON.
It is recommended to set the goal for both channels to the MVC of the healthy side
to provide the most accurate percentage deficit. Failure of any of these steps will
impact test results.
15)
Tap Track and select a timestamped session to see actual EMG activity compared
to goal for a saved session. Tap an individual graph to see average maximum
contraction for that session. You can also delete individual sessions using your
mobile devi
ce’s standard delete gesture.
Note
: mTrigger data is stored locally and only identified by timestamp on the mobile
device on which the session was recorded. It is not remotely accessible or backed
up to any additional location.