![MSI Port-A-Weigh MSI-4260 User Manual Download Page 26](http://html1.mh-extra.com/html/msi/port-a-weigh-msi-4260/port-a-weigh-msi-4260_user-manual_1848750026.webp)
M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L
Page 26
MSI-4260
Port-A-Weigh
•
User Guide
3) Pick up a test weight of (or set the hydraulic tension to) at least 25% of capacity. The
weight is indicated on the numeric digits. The display will
fl
ash "LOAd" if the weight is
under 25%.
WARNING:
Fine Calibration will not function unless the scale is loaded at 25% of capacity
or more.
3) Press
[SELECT/F1]
to cause the displayed reading to move up slightly. Press the button
[CLEAR/ZERO]
to cause the reading to move down. Each press of the
[SELECT/F1]
or
[CLEAR/
ZERO]
button causes the calibration to shift by 1/4 displayed count. When the displayed
reading is acceptable, press
[ENTER/TARE]
.
Final
Fine Calibration is complete, press
[CLEAR/ZERO]
to return to normal scale operation, or press
[SELECT/F1]
to choose another Calibration Setup function.
CALIBRATION WITH CALIBRATION OFFSET RESISTOR (CAL-r)
Function
Calibration with the Cal Resistor (CAL-r) is for emergency calibration only when test
weights or a calibration system are not available.
Action
1) Press
[SELECT/F1]
to scroll to the "CAL-r" message. Press
[ENTER/TARE]
.
2) Display
fl
ashes "UnLd". Unload the scale and press
[ENTER/TARE]
. The scale will display
"CALC" (for calculating).
3) Display shows "LOAd" for 2 seconds then
fl
ashes last set CAL-r value. Input the CAL-r
value printed on the load cell Cal Sheet. After the last number has been entered, press
[ENTER/TARE]
. The display will show "CAL’d".
4) Display
fl
ashes "UnLd". Unload the scale and press
[ENTER/TARE]
. Afterwards CAL-r
will be displayed until the value is determined. Then the CAL-r value is displayed for 10
seconds, or until any button is pressed.
Final
Calibration is complete and the display shows "StorE" to signify that the calibration param-
eters are being saved in backup memory. The calibration menu returns. Press
[CLEAR/ZERO]
to return to normal operation or press
[SELECT/F1]
to scroll to a different Calibration Setup
operation.
Note:
When Calibration is
fi
nished, seal the Calibration Port on the front panel of the scale.
STANDARDS (StAnd)
Function
Facilitates calibration to your required standard:
IndUS
Industrial
nISt
National Institute of Science and Technology (NTEP setting)
EuroP Metric/OIML
1Unit
One Unit only (for custom scales)
WARNING
WARNING