CALIBRATION SETUP
Function
Calibration Setup is used to initiate calibration or to set seldom changed factory settings.
There are two types of calibration:
1) Initial Calibration;
2) Subsequent Calibration
Calibration must be accomplished by a quali
fi
ed scale technician trained in Certi
fi
ed Calibra-
tion Standards. To initiate calibration of a Port-A-Weigh, the technician is required to have an
accurate test weight system of adequate capacity and, in the case of a Legal For Trade Port-
A-Weigh, the test weight system must be certi
fi
ed by the appropriate regulatory agency.
Important Note:
At any point (except when entering calibration weight values), the calibra-
tion procedure can be halted by turning POWER off.
Action
1) Turn on the scale by pressing the
[POWER]
button.
2) Remove the seal port on the right side of the scale’s front casting.
3) Press the Cal switch button in the hole with your
fi
nger (see photos on page 21). The
display will show "C-SEt" for 2 seconds.
4) The Calibration Setup Menu appears. Press the
[SELECT/F1]
button to scroll through the
various Calibration Setup menu choices.
5) Once the desired operation is displayed, press
[ENTER/TARE]
.
Final
To return to normal operation either scroll with the
[SELECT/F1]
button to exit and press
[ENTER/TARE]
, or press
[CLEAR/ZERO]
.
QUICK CALIBRATION SETUP GUIDE
Top Menu
Selection Display
Display De
fi
nition
CAL
Calibration
F-CAL
Fine
Calibration
CAL-r
Calibration with Offset Calibration Resistor
StAnd
Standards
IndUS
Industry
nISt
National Institute of Standards & Technology
EuroP
Metric (kg only)/OIML
1 Unit
One Unit Only
r-Ctl
Remote
Control
On
OFF
AZM
Automatic ZERO Maintenance
On
Off