TARE
T
TARE
T
TARE
T
ZERO
O
sto
ZERO
O
sto
USER
F
USER
F
If you are already in the Cal menu
skip to step 2
1)
Remove the seal screw (cut
lead wire seal if present). With
a small screwdriver press the
Cal Switch behind the seal
screw.
The CAL Setup Menu
appears.
2)
Press the
USER
key to
scroll to ”Setup” .
3)
Press
TARE
to enter the Cal
Setup Menu.
4)
Press
TARE
to enter the
Standard Menu. The current
standard setting is displayed.
5)
Scroll to the desired standard
using the
USER
key. .
6)
Press
TARE
to set the
standard. The display goes to
the next item in the CAL Setup
Menu.
7)
Either press
ZERO
twice to
exit Setup and store all
changes, or continue to
another Setup Menu item
using the
USER
Key.
&$/
LFT Standard Setup
VHWYS
6WDQG
LQGXV
XQLW
$YWR
Next Cal Setup Menu Item
Store and return to
weight display
&D/
6WRUH
Calibrate Menu
Remove
Hex Seal
Screw
Cal Switch
Access
Press Switch
behind the hole
with a small
Screwdriver
40
MSI-4260 Port-A-Weigh Operator’s Manual
Figure 5-9. LFT Standard Setup
5.5 Auto Zero Maintenance
The
MSI-4260
employs an autozeroing maintenance mechanism to adjust the zero
reading to the Center-of-Zero (COZ) COZ is defined as the weight reading is within
1/4 ‘d’ of zero. AZM continuously adjusts zero to maintain COZ. It is recommended
that AZM is on to maintain the highest accuracy. However, there are circumstances
when it should be turned off. This can happen when minor variations of weight occur
while picking up scale attachments and the variations fall within the AZM capture
window. The AZM capture window (usually 1 ‘d’) and capture time (usually 8
seconds) can be adjusted by MSI to meet custom requirements. The settings of AZM
are dictated in Legal-for-Trade Standards and cannot be adjusted.