Cal Standard
Auto Zero
Maintenance
Calibrate with Weights
Calibrate with Constant
Industrial Setting
(default)
Handbook 44, NTEP
R-76, OIML
Industrial, restricted to
1 Weight Unit
AZM On
AZM Off
Remove
Hex Seal
Screw
Cal Switch
Access
Press Switch
behind the hole
with a small
Screwdriver
USER
F
Calibration
39
Figure 5-8. Calibration Setup Menu Map
5.4.1
Standard
There are 4 selections in the standards menu.
1. Industrial (indu5) – This is the most common setting for the 4260. With the
Industrial standard, you have full range zero, access to units switching, filters,
and peak hold.
2. Handbook 44 (Hb-44) – Sets the scale to enable only approved features per the
NTEP HB-44 rules and regulations. Access is denied to Peak Hold, and the zero
range may be limited. The Filter menu is moved to the Cal Setup Menu, so
filters are only accessible through the Cal Seal.
3. R-76 (r-76) – Sets the scale to enable only approved features per OIML R-76.
Only kg weight units are available. The zero range is limited to 5% (-2 +3%
relative to Calibrate zero). Net/Gross function is temporary. once Net weight is
established, pushing an F key set for Net/Gross will cause a maximum 5 second
display of the Gross weight. You must clear the Tare to display Gross weight
constantly. Other metrological aspects are changed to meet R-76 requirements.
4. One Unit (1unit) – The one unit Standard is exactly the same as Industrial,
except units switching is inhibited. This is useful for Metric only countries.
Another use of the One Unit standard is to allow the scale to be calibrated in
units other than lb or kg, since conversions are eliminated.
Contact MSI for more information on the Standards settings.