4. ROTORS AND ACCESSORIES
4.1 Installation Of Rotor
Before fitting a rotor to the centrifuge ensure that the drive shaft is lightly smeared
with petroleum jelly. Check that the rotor is clean - particularly pockets in angle
rotors.
Locate the rotor on the drive shaft and, ensuring that the correct spindle nut for the
rotor in use is employed (see sections 0 & 4.5), engage and tighten the spindle nut
finger tight.
4.2 Removal Of Rotor
Press Power switch off. Open the centrifuge lid. Remove all rotor accessories, i.e.
buckets, trunnions etc. Loosen the spindle nut a few turns, grip the rotor as
illustrated in Figure 4 and pull upwards with the fingers whilst gently rocking the
rotor and applying downward pressure with the thumbs, to release it from the drive
shaft. Remove the nut and lift the rotor free.
Note: Do not attempt to loosen the rotor by tapping on the centrifuge motor
spindle.
Figure 4.
Removing Rotor from Drive Shaft