INSTALLATION
12
3. INSTALLATION
IMPORTANT: The rotor must not be fitted when moving or during transit.
3.1 Preparation
Remove the centrifuge from its packing and place it on a rigid, level surface.
Check that the available power supply corresponds to that stated on the voltage
plate fixed to the rear panel of the centrifuge.
NOTE: (i) An electrical interlock prevents the lid being opened unless the
centrifuge is connected to the power supply.
(ii) It is important that, at least, 100mm clearance is allowed at the
sides and rear of the centrifuge when placed in the final operating
position.
NOTE: Laboratory management procedures should require that no person or
any hazardous materials are within a 300mm boundary while the
centrifuge is operating.
3.2 Connecting To 3 Pin Plug
Connect the 3-core cable to a 3-pin plug fitted with a 5 Amp fuse as follows:
WARNING! This equipment must be earthed.
Brown wire to Live (L) Terminal
Blue wire to Neutral (N) Terminal
Yellow/Green wire to Earth (E) Terminal
Connect the centrifuge to the power supply and press the Power switch on. The
Power On and the Lid Unlocked indicators will come on. Push the lid lock actuator
to the right and open the lid.
NOTE: BS EN 61010-2-020 ‘Particular Requirements for Centrifuges’ states that
a remote switch, preferably adjacent to the room exit is a requirement
in case of unit malfunction.