User manual
DELTA
smart
MRU GmbH, D-74172 Neckarsulm
39 / 50
9
Extras
Settings
The
DELTA
smart
is delivered with a pre-configured firmware which co-
vers most of the needed applications. This unit can however be individ-
ually configured to meet anyone’s needs.
9.1.
Service calibration menu
The service and calibration menu is pin code protected to assure the
miss-use of unauthorized personnel.
When entering a wrong pin code, you will automatically be returned the
extras menu.
Contact MRU or a MRU authorized dealer to get the pin code
(www.mru.eu).
9.2.
Factory settings
All parameters will be reset to factory settings:
Parameters for measurement that have been modified by the user will
be set to factory settings again.
Settings:
LCD brightness (%)50
Helping hints ON
Keyboard beep ON
All changes that have been made by the user (Combustion test) screen
settings will be re-set to factory settings.
9.3.
Service values
In case of a defect sensor, which will be displayed after zeroing, the ser-
vice values of all installed sensors and other parameters can be viewed
in this menu.
When you experience a defect of a sensor you will most likely call in at
MRU and they will ask you for these service values to identify the defect
sensor or other defect parts. Each sensor will display a service value as
well as other parts installed inside the analyzer.