6.6.
Controlling condensate separator (water trap)
The condensate separator must be emptied before and after each
measurement, you will get a reminder to do that when switching off the
analyzer.
The condensate separator must be checked before, during and after
each measurement.
Check if the condensate separator is
empty
and if the pleated filter is
white (clean).
CAUTION
Condensate is acidic
The condensate from the container can be slight acidic
and can cause chemical burn.
►
Immediately clean with plenty of water once you
have encountered acid
Here is how to empty the condensate separator:
•
Move the hoses from the upper bracket to the side.
•
Unscrew upper plug (counter clock wise) then lift it.
•
Empty the container with caution.
•
The condensate separator can easily be disassembled / assembled to
remove the filter and to let the condensate separator dry.
•
Both plugs have gaskets, which must be present, undamaged and in
the correct position when you assemble the condensate separator
again.
6.7.
Connections and tightness
•
Check if all connections are connected properly.
•
Check the complete probe, sampling line and condensate separator
for leaks / leak tightness.
The
DELTA
smart
offers an automated test to check if the complete sys-
tem is leak tight.
See also Chapter 9.4 Analyzer and accessories leak test, Page 40.