Setting Zeropoint
Leave the analyser in good fresh air for five minutes, if possible, without
changing the temperature. Now you can take a zero point. You can re-
peat a zeroing as often as you like.
► Select
„
Set Zero
“
.
Press
„
OK
“.
A window appears.
Zero point is set automatically.
►
Press „OK“.
The menu Extras appears.
Setting alarm threshold
►
Select „
Alarm
“.
►
Press
„OK“.
►
Set the desired alarm threshold.
►
Press
„OK“.
The alarm threshold is saved.
The menu Extras appears.
NOTE
Pay attention to the unit for which you set an alarm
threshold for the interchangeable sensor CO400.
You can select CO ppm or CO mg/m³.
►
If necessary, change the unit.