If the experience and instruction for use the instructions in the manual does not
match or not relates to the situation, please timely contact with
the manufacturers, to upgrade or update.
Manual save to keep ventilation drying, avoid high humidity and temperature.
3.Summarize
The product use scope
Applicable to medical and health services, scientific research, agricultural and
other units, to the medical equipment, dressings, glassware, solution culture
medium to carry out sterilization, also is a necessary test equipment food
factory, drinking water plant to do QS, HACCP certification.
Technical parameters
Design pressure: 0.28MPa
Rated working pressure: 0.22MPa
Working temperature: 134 degrees
Source: Deionized water (pure water, distilled water)
Volume: 60/80/100L voltage: 220V
Hygiene license number: (Lu) Xiao Wei card characters (2014) 504th
Design service life: 5 years
Minimum equipment sales unit: 1
Attention
1. Please open the door after discharging all the steam. Do not close to the
sterilizer at the same time
2. The chamber is still keep at high temperature after the sterilizing, please pay
attention to insulate the chamber in case of scald.
3. Monitoring method
Methods to monitor sterilization effect including adopting
temperature-inspecting sterilization test paper, biological indicator
4. Please avoid the safety valve towards people or other equipment in case of
scalding by steam.
4.Equipment principle and range of application
STE-V series full-automatic vertical autoclave has a microprocessor control
system, it can sterilizing at constant temperature steadily. The equipment
use
LED display, touch button and easy one button operation with a high
control
accuracy in different ambient temperature, heating power and
voltage
fluctuation
Model STE-L series autoclave use immersion heating with a high heating