First measurement is at cryogenic temperatures (10-30) (can be replaced with
an analog temperature sensor) to record the actual temperature T1,and
temperature displayed is D1.
Then measurement at the high temperature (100-130) (can be replaced with an
analog temperature sensor) to record the actual temperature T2, D2 shows the
temperature displayed
Result (T2-T1) / (D2-D1) multiplies parameter K displayed on tEFA menu is
obtained a new parameter
After modifying the parameters in tEFA , then according to the temperature
difference between D3(the temperature displayed on the temperature sensor )
and T2 to modify the parameters in tErF , if the displayed temperature is low,
then the value in tErF plus the temperature difference, if the displayed
temperature is high, the value in the tErF subtracts the difference.
6.Operation Run
Sterilizer operating procedures, include sterilization preparat ion, sterilization
items loading, sterilization, sterile unloading procedure.
6.1. Sterilization preparation
(1)Cleaning: article should be thoroughly cleaned before sterilization, avoiding
the presence of blood and other impurities, as these substances would harm
sterilizing articles and sterilizer. After washing , items should be timely dried and
packaged.
(2) Packaging: For packaging materials, please use beneficial internal air exhaust
and steam penetration packaging materials, strictly complying to "Disinfection
technical specifications" and the relevant national standards. The following
points may be beneficial to your sterilization effect:
①
plate, pots, bowls and other containers, as a single package, and the lid is
opened.
②
surgical instruments should be placed in a basket or perforated plate for
supporting the package.
③
items must be exposed , when the stack release, utensils should be used
between moisture cloth, gauze or medical absorbent paper separated.